nEt
Programming of the net weighing capacity in kg (approximate).
A value lower than 1/10 of
CAPAC
is not accepted.
dEAd L
Programming the fixed tare value in kg (approximate)
dSPdiV
Divisions value
The division value is expressed in kg, selectable between 0.0001 kg and 50 kg.
The ratio between the maximum system capacity and the division value is the system
resolution (number of divisions).
Following the modification of the system capacity, a division value is automatically select-
ed at the best of the 10,000 divisions.
Following the modification of the division value, if the maximum capacity is not changed,
the weight calibration is automatically corrected.
SiGnAl
Display of the input signal in mV / V (from -0.50 to 3.50 mV / V).
AnAloG
Programming of the weight corresponding to the full scale of the analog output.
(Standard: scale capacity)
SEnSit
Load cell sensitivity
Set the value corresponding to the average of the sensitivities at the nominal capacity of
the load cells, in mV / V. Values between 0.5 and 4 mV / V are accepted. If no value is
programmed, 2mV / V is assumed.
Following the modification of the sensitivity value, the theoretical weight calibration is per-
formed.