20
Weight calibration and linearization
During the calibration phase, the display shows the weight intermittently.
Zero calibra-
tion
Carry out the operation with the scale unloaded but complete with the tare, at stabilized
weight. The displayed weight must be reset. It is possible to repeat this operation several
times.
Zero and full scale calibrations can be canceled.
+
cancels the zero calibration
+
cancels the full scale calibration.
Full scale cali-
bration
Before carrying out the operation, load the sample weight on the scale and wait for stabili-
zation; the display shows the measured value to be calibrated.
If the set value is higher than the resolution offered by the instrument, it is not accepted
and the display shows an error message for a few seconds.
It is always possible to repeat the calibration operations
Se si spegne lo strumento senza uscire dal menu di set-up non vengono memorizzate le
programmazioni eseguite.
!
When programming the sample weight, values higher than the full scale, or lower than the previous point,
or with an unstable weight are not accepted. If the entered value is accepted, the next point is proposed,
otherwise still the same. The linearization points are automatically zeroed every time a value of the theoreti-
cal calibration is changed or if the 0 or the full scale are reset.
Linearization procedure
Keep
pressed
3 sec.
CAlibr ProGr PArAM in-out SEriAl AnAloG ConFiG
StorE
Exit menu
CAL
Zero calibration
Set sample weight
F.S. calibration
CAL
3 sec.
Lin P1
Set sample weight 1
st
step
Lin P2
Lin P9