Pagina –
page
–
Seite
17
Dosing operation on discharge
When the unloaded net weight reaches the SET - PRESET value, the FAST unloading output 1 will be
opened. When the unloaded net weight reaches the SET - TAIL value, the SLOW unloading outlet 2 will be
opened, the stable weight will be waited for and after 3 seconds, the autotare will be canceled and the
weight present on the scale will be displayed. If the full scale of the system is exceeded, the dosage is inter-
rupted.
Dosing operation at loading without zeroing
The instrument allows to dose a product with two extraction speeds. Close the START contact or press the
FUN key to start the dosage:
If the full scale of the system is exceeded, the dosage is interrupted.
During the dosage, the letter D will appear on the left of the display and the gross weight. When the net
weight reaches the SET - PRESET value, the FAST load output 1 will be opened. When the net weight
reaches the SET - TAIL value, the SLOW loading output 2 will be opened, the stable weight will be waited
for and after 3 seconds the weight present on the scale will be displayed without the letter D.
To stop the dosage at any time, close the STOP input or press the FUN key. At start the two outputs of
SLOW load and FAST load are closed simultaneously.
If the SET load value added to the gross weight is greater than the maximum weight set, the message
OVEr
will be displayed for 3 seconds and the dosage will not start.
If the SET load value is lower than the CODA, the
ErCodA
message will be displayed for 3 seconds and the
dosage will not start.
If the SET load value previously set is zero, the message
noSEt
will be displayed for 3 seconds and the
dosage will not start.
The tool allows you to dose a product with two extraction speeds by performing the autotare at each cycle
start. Close the START contact or press the FUN key to start the dosage:
If the SET discharge value previously set is zero, the message
noSEt
will be displayed for 3 seconds and
the dosage will not start.
If the SET discharge value is lower than the CODA, the
ErCodA
message will be displayed for 3 seconds
and the dosage will not start.
If the SET discharge value is greater than the gross weight on the scale at start, the message
noProd
is
displayed for 3 seconds and the dosage does not start.
To stop the dosage at any time, close the STOP input or press the FUN key.
At the start the autotare of the weight present in the scale is performed, the instrument automatically switch-
es to the net weight display and the two outputs of SLOW load and FAST load are closed simultaneously.
During the dosage, the letter D will appear on the left of the display and the net weight.