47
Weight display, reset and autotare
On power up, the display shows the current net weight.
Switching between net weight and gross weight display
1
5
0
5
4
2
3
5
Press the
key to switch the display from net weight to gross weight and
vice versa. The displayed value is indicated by the upper LED (lit: net
weight). If the tare is not entered, the net weight is equal to the gross weight.
This operation is saved on shutdown. In case of negative weight, the minus
sign is displayed before the most significant digit. In the case of a negative
weight greater than 9999, the minus sign is displayed alternating with the
most significant digit.
Switching between numeric display / bar-graph of the gross weight
Press the
key to switch from the numerical display of the weight to the
graphic representation of the gross weight and vice versa. The resolution is
limited to 15 divisions and therefore each segment represents 1/15 of the full
scale. The bar-graph display is indicated by both flashing LEDs.
=
=
=
4
2
3
5
Restore the zero (semi-automatic zero) in the gross weight display
=
=
=
4
2
3
5
3 sec.
This operation is performed to correct small zero movements of the scale.
To carry out the zeroing function it is necessary to switch the display to the
gross weight (NET LED off). The gross weight reset command is not per-
formed in the following situations:
Unstable weight (the weight does not stabilize within 3 seconds from the
reset command)
Gross weight, greater than the original zero calibration, greater (positive
or negative) than the set
0bAnd
value. In this case it is necessary to
enter the calibration menu and perform the calibration
If there is an autotare, this is automatically canceled. The semi-automatic zero operation is canceled when
the instrument is turned off.
Auto weighted tare (Autotara) in net weight display
=
=
=
4
2
3
5
3 sec.
If the autotare with gross weight = 0 is performed, the possible tare value is canceled. The entered tare val-
ue is not saved when the instrument is turned off. If a tare is inserted, this is signaled by the second LED on
(on: tare inserted).
To carry out the autotare function it is necessary to switch the display to the
net weight. (NET led on). The autotare command is not performed in the
following conditions:
Unstable weight (weight does not stabilize within 3 seconds)
Gross weight negative
Gross weight exceeds the maximum capacity of the system
Содержание RQ
Страница 1: ...R QS o w a r e P W 0 3 1 4 0 3 V e r s i o n e 1 0 Manuale Manual Handbuch...
Страница 16: ...Pagina page Seite 14 Appunti Notes Notizen...
Страница 46: ...Pagina page Seite 44 Appunti Notes Notizen...
Страница 76: ...Pagina page Seite 74 Appunti Notes Notizen...
Страница 97: ...Pagina page Seite 95 Appunti Notes Notizen...
Страница 98: ...Pagina page Seite 96 Appunti Notes Notizen...
Страница 100: ......