Return to main menu
A
d
d
r
Select value
M
o
d
E
Select value
d
E
l
A
y
Set value
Serial communication address COM1
Programming of the address used in the weight transmission
protocols and for the MODBUS protocol, a value that can be set
from 0 to 99.
Weight data transmitted COM1
Selection of the value transmitted with continuous, manual and
automatic protocols (see relative paragraph).
nEt GroSS PEAk
Slave and ModBus protocol COM1 response delay
Value expressed in milliseconds, it represents the delay with
which the instrument sends the response to the master's re-
quest, if the latter is not able to receive immediately at the end of
the transmission.
i
p
Set value
Set IP address
Sequential setting of the 4 characters of the IP address. Ex.:
if IP address = 10.0.0.201 set 10 in IP1, 0 in IP2, 0 in
IP3 and 204 in IP4.
S
u
b
n
t
Set value
ProfiNet subnet mask
Programming of the subnet mask used for the ProfiNet protocol.
The 4 bytes that make up the subnet mask must be pro-
grammed in sequence (as illustrated on the previous page for
programming the IP address).
IP and SUBNET are only visible in the Ethernet IP version.
With PROFINET they are assigned by the master!
Содержание RQ
Страница 1: ...R QS o w a r e P W 0 3 1 4 0 3 V e r s i o n e 1 0 Manuale Manual Handbuch...
Страница 16: ...Pagina page Seite 14 Appunti Notes Notizen...
Страница 46: ...Pagina page Seite 44 Appunti Notes Notizen...
Страница 76: ...Pagina page Seite 74 Appunti Notes Notizen...
Страница 97: ...Pagina page Seite 95 Appunti Notes Notizen...
Страница 98: ...Pagina page Seite 96 Appunti Notes Notizen...
Страница 100: ......