ERROR MESSAGES
There are indications for the main anomalies. The display shows the messages with the operative
situations and the lights
WARNING
and
ALARM
.
Remember that, for a correct operation, the system must have a capacity of 5 litres/min. and that the
water must have a temperature of 20°C max and a pressure of at least 2 bar.
1)
In case of NO water or low water pressure:
Error !
Water press. low
THE INSTRUMENT DOES NOT START (Fig. 8)
2)
After 10 seconds the display shows the message
No water
THE INSTRUMENT DOES NOT START
If the error situation continues the distillation will be cancelled and the program will wait to confirm
the reading of the message by operator.
No water
Ok
CONFIRM THE READING OF THE MESSAGE.
3)
In case of cooling water too hot the yellow light flashes. Push the START key to start a distillation.
The instrument will wait 10 seconds and if the error situation remains the operator will select if to
continue or cancel the distillation.
Water too hot
Go on Cancel
CONFIRM OR CANCEL THE DISTILLATION
(Fig. 9)
ATTENTION!
The temperature of the cooling water upper 20 °C compromises the result of the distillation
whether in alcohol or Volatile acidity mode.
Figure 8
Figure 9
Many situations of error have the indication of the error message until the restored normal situation.
4)
In case of drainage open the following message will appear:
Attention!
Drainage open
THE DISTILLATION DON’T START.
THE INSTRUMENT WAITS FOR THE CLOSURE
OF THE DRAINAGE (Fig. 10)
5)
In case of open plexiglass the following message will appear:
Attention!
Plexiglass open
THE DISTILLATION DOES NOT START.
THE INSTRUMENT WAIT FOR THE CLOSURE
OF THE PLEXIGLASS (Fig. 11)