PMsense
- 11 -
V1.0
Command
Reply
Description
CPSn
&|
Sets the type of measurement averaging for the contin-
uous measurement mode:
Average over a 10 seconds interval, updated every
second if n=0
Average over a 60 seconds interval, updated every
10 seconds if n=1
Average over a 15 minutes interval, updated every
minute if n=2
Default
: Average over a 60 seconds interval, updated
every 10 seconds (n=1)
RPS
& n|
Reads the type of measurement averaging for the con-
tinuous measurement mode.
Analog outputs (only PM[B]sense-A):
Command
Reply
Description
CA1On
&|
Enable/disable the offset of the analog output 1:
Offset disabled if n=0 (0…20 mA or 0…10 V)
Offset enabled if n=1 (4…20 mA or 2…10 V)
Default
: Offset enabled (n=1) if the output is current,
offset disabled (n=0) if the output is voltage (optional)
RA1O
& n|
Reads the setting of the offset for the analog output 1.
CA1SOn
&|
Sets the direct or inverse correspondence between analog
output 1 and associated physical quantity:
4 mA/0 V
Min. quantity, 20 mA/10 V
Max.
quantity if n=0
20 mA/10 V
Min. quantity, 4 mA/0 V
Max.
quantity if n=1
Default
: Direct correspondence (n=0)
RA1SO
& n|
Reads the type of correspondace (direct or inverse) be-
tween analog output 1 and associated physical quantity.
CA1Tn
&|
Associates the analog output 1 to:
PM1.0 if n=0
PM2.5 if n=1
PM10 if n=2
CO
2
if n=12 (only PMBsense-A)
Default
: PM2.5 (n=1)
Note: the PM measurement is updated according to the op-
erating mode set with the CPLS, CPLP and CPS commands.
RA1T
& n|
Reads the physical quantity associated to analog output 1.
CA1Ln
&|
Sets n as the minimum value of the measuring range of
the physical quantity associated to the analog output 1.
If the output is associated to PM, the value must be ex-
pressed as a number of tenths (e.g. n=5 to indicate 0.5
µg/m
3
). If the output is associated to CO
2
, the value must
be expressed as a number of units (e.g. n=200 to indicate
200 ppm).
Default
: 0
RA1L
& n|
Reads the minimum value of the measuring range of the
physical quantity associated to the analog output 1.
Содержание Delta OHM PMBsense-A