![GHM Delta OHM PMBsense-A Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/ghm/delta-ohm-pmbsense-a/delta-ohm-pmbsense-a_operating-manual_2220397017.webp)
PMsense
- 17 -
V1.0
39
38
Internal board temperature (x10).
16-bit Integer
41
40
Instrument firmware revision. The most significant byte
indicates the major revision; the less significant byte
indicates the minor revision.
16-bit Integer
42
41
Number of MODBUS communication error.
16-bit Integer
(*)
The measure is a 32-bit integer value. Two consecutive registers must be accessed to read
the value. The register with lower address contains the most significant bits.
Coils:
Number Address
Description
Format
1
0
Set 1 to restore the factory configuration. Bit zeroing
is automatic.
Bit
2
1
Enable configuration change:
0=no (default), 1=yes.
The changes to
Coils
and
Holding Registers
will be ac-
cepted only if this register is set to 1.
Bit
3
2
Sets waiting time after transmission with MODBUS-RTU
protocol:
0=immediate reception (default);
1=waiting 3.5 characters.
Bit
4
3
Enable/disable the offset of the analog output 1:
0=offset disabled (0…20 mA or 0…10 V, default if the
output is voltage);
1=offset enabled (4…20 mA or 2…10 V, default if the
output is current).
Bit
5
4
Sets the direct or inverse correspondence between ana-
log output 1 and associated physical quantity:
0=4 mA/0 V
Min. quantity, 20 mA/10 V
Max.
quantity (default);
1=20 mA/10 V
Min. quantity, 4 mA/0 V
Max.
quantity.
Bit
6
5
Enable/disable the offset of the analog output 2:
0=offset disabled (0…20 mA or 0…10 V, default if the
output is voltage);
1=offset enabled (4…20 mA or 2…10 V, default if the
output is current).
Bit
7
6
Sets the direct or inverse correspondence between ana-
log output 2 and associated physical quantity:
0=4 mA/0 V
Min. quantity, 20 mA/10 V
Max.
quantity (default);
1=20 mA/10 V
Min. quantity, 4 mA/0 V
Max.
quantity.
Bit
Содержание Delta OHM PMBsense-A
Страница 22: ...PMsense 22 V1 0 NOTES...
Страница 23: ......