![GHM Delta OHM PMBsense-A Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/ghm/delta-ohm-pmbsense-a/delta-ohm-pmbsense-a_operating-manual_2220397016.webp)
PMsense
- 16 -
V1.0
11
10
PM2.5 in µg/m
3
(x10).
Measurement averaged over 10 s and updated every
second. Available in continuous measurement mode.
16-bit Integer
12
11
PM10 in µg/m
3
(x10).
Measurement averaged over 10 s and updated every
second. Available in continuous measurement mode.
16-bit Integer
13
12
PM1.0 in N° of particles/ml.
Measurement averaged over 60 s and updated every
10 s. Available in continuous measurement mode.
16-bit Integer
14
13
PM2.5 in N° of particles/ml.
Measurement averaged over 60 s and updated every
10 s. Available in continuous measurement mode.
16-bit Integer
15
14
PM10 in N° of particles/ml.
Measurement averaged over 60 s and updated every
10 s. Available in continuous measurement mode.
16-bit Integer
16
15
PM1.0 in µg/m
3
(x10).
Measurement averaged over 60 s and updated every
10 s. Available in continuous measurement mode.
16-bit Integer
17
16
PM2.5 in µg/m
3
(x10).
Measurement averaged over 60 s and updated every
10 s. Available in continuous measurement mode.
16-bit Integer
18
17
PM10 in µg/m
3
(x10).
Measurement averaged over 60 s and updated every
10 s. Available in continuous measurement mode.
16-bit Integer
19
18
PM1.0 in N° of particles/ml.
Measurement averaged over 15 min and updated every
minute. Available in continuous measurement mode.
16-bit Integer
20
19
PM2.5 in N° of particles/ml.
Measurement averaged over 15 min and updated every
minute. Available in continuous measurement mode.
16-bit Integer
21
20
PM10 in N° of particles/ml.
Measurement averaged over 15 min and updated every
minute. Available in continuous measurement mode.
16-bit Integer
22
21
PM1.0 in µg/m
3
(x10).
Measurement averaged over 15 min and updated every
minute. Available in continuous measurement mode.
16-bit Integer
23
22
PM2.5 in µg/m
3
(x10).
Measurement averaged over 15 min and updated every
minute. Available in continuous measurement mode.
16-bit Integer
24
23
PM10 in µg/m
3
(x10).
Measurement averaged over 15 min and updated every
minute. Available in continuous measurement mode.
16-bit Integer
27
26
PM measurement error: 0=no, 1=yes.
Intero 16 bit
29
28
CO
2
in ppm.
Intero 16 bit
34,35
33,34
Atmospheric pressure in Pa
(*)
Internal sensor for CO
2
measurement compensation
Intero 16 bit
36
35
Atmospheric pressure in hPa (x10)
Internal sensor for CO
2
measurement compensation
Intero 16 bit
38
37
Power supply voltage in Volt (x10).
16-bit Integer