background image

8850-2 Conductivity/Resistivity Transmitter 

Hysteresis 

Time 

Low Setpoint 

Process 

Relay energized

Relay de-energized

Hysteresis 

Time 

High Setpoint

Process

3.3 Relay Output

The relays can be used as a switch that responds when the 
process value moves above or below a setpoint, or to generate a 
pulse at a rate proportional to the process value.

• Low:

Relay triggers when process is less than the setpoint.
The relay will de-energize when the process moves above the 
setpoint plus the hysteresis value.

• High:

Relay triggers when process is greater than the setpoint.
The relay will de-energize when the process moves below the 
setpoint plus the hysteresis value.

Signet 

Conductivity/Resistivity

Transmitter

ENTER

62.50 uS/cm

25.0 C

VIEW menu

•  During normal operation, ProcessPro displays the VIEW menu.
•  When editing the CALIBRATE or OPTIONS menus, ProcessPro will return to the VIEW menu if no 

activity occurs for 10 minutes.

•  To select a VIEW display, press the 

 or 

 keys. The selections will scroll in a continuous loop.

•  Changing the VIEW display does not interrupt system operations.
•  No key code is necessary to change display selection.
•  Output settings cannot be edited from the VIEW menu.

• Proportional Pulsing

The relay will generate a 100 mS pulse at the rate de

fi

 ned by 

settings in the CALIBRATE menu.

In the example below:
•  The output will be 0 pulses/min. when value is less than 5.
•  The output will be 50 pulses/min. when value is 7.5.
•  The output will be 100 when value is greater than 10.

The starting point, endpoint and maximum pulse rate are select   -
able in the CALIBRATE menu.

10

5

0 pulses

Relay Output Rate:

0 to 100 Pulses/min.

endpoint

100 pulses

Starting point

Pulse rate

View Menu

Display

Description

1234.5 

μ

S/cm

12.6  °C

Monitor the Conductivity and the Temperature input from the sensor.
This is the Permanent display.

All of the displays below are temporary. After 10 minutes the display returns to the permanent display.

Loop Output

14.16 mA

Monitor the 4 to 20 mA Loop output.

Last CAL:

02-10-09

Monitor date for scheduled maintenance or date of last calibration.
(See description in Calibrate Menu.)

Содержание Signet ProcessPro 8850-2

Страница 1: ...nder ISO 9001 and ISO 14001 CAUTION Remove power to unit before wiring input and output connections Follow instructions carefully to avoid personal injury Contents 1 Installation 2 Specifications 3 El...

Страница 2: ...val Procedure 1 Press the orange terminal lever downward with a small screwdriver to open terminal jaws 2 When fully open remove wire from terminal Wiring Tips Do not route sensor cable in conduit con...

Страница 3: ...Loop AUX Power AUX Power Transmitter 2 Terminals PLC Terminals Loop Input 4 20 mA Transmitter Terminals System Pwr Loop System Pwr Loop AUX Power AUX Power Power Supply 12 24 VDC Power Supply 12 24 VD...

Страница 4: ...keys The selections will scroll in a continuous loop Changing the VIEW display does not interrupt system operations No key code is necessary to change display selection Output settings cannot be edite...

Страница 5: ...2 The Key Code is keys in sequence After entering the Key Code the display will show the first item in the selected menu Step 3 Scroll menu with or arrow keys Step 4 Press key to select menu item to b...

Страница 6: ...ut Loop Range S 0 0000 100 00 Select the minimum and maximum values for the 4 to 20 mA current loop output Be sure to modify this setting if you change the Cond Units Relay1 Source Cond Select Conduct...

Страница 7: ...able display and output response Temp Display C Select temperature units C or F Temperature Comp 2 00 Set the percent change in Conductivity caused by a 1 C change in temperature May be from 0 00 to 1...

Страница 8: ...t value If the instrument does not display the correct value Temperature 0 5 C Conductivity 2 of reading service is required A Accuracy Verification with Precision Resistors Electronic Calibration 1 S...

Страница 9: ...ayed temperature and conductivity values in the spaces provided Displayed temperature T1 _______ C Displayed conductivity C1 _______ S 3 Cool the sample solution close to the minimum process temperatu...

Страница 10: ...condition if your process operates at or near the limits of the electrode range Check sensor for correct range Check Decimal setting in OPTIONS menu Check Calibrate menu settings for incompatible SOUR...

Страница 11: ...11 8850 2 Conductivity Resistivity Transmitter Notes...

Страница 12: ...085 CR Sanitary 0 01 SS 1 to 1 1 2 in 3 2819 S1C 159 000 087 CR Sanitary 0 01 SS 1 to 1 1 2 in Certified 3 2819 S2 159 000 086 CR Sanitary 0 01 SS 2 in 3 2819 S2C 159 000 088 CR Sanitary 0 01 SS 2 in...

Отзывы: