background image

10 

8850-2 Conductivity/Resistivity Transmitter 

Troubleshooting

If a Current Loop is locked at 3.6 mA, the problem is related to the temperature circuit:
This occurs only if the 8850 detects a resistance from the temperature sensor that is less than 250 

Ω

 or greater than 2800 

Ω

.

•  Check the sensor wiring for open/short or poor connections on white (TEMP IN) and black (ISO GND)wires.
•  The PT1000 temperature device in the sensor is defective.
•  The transmitter is defective.

Auxiliary Power (terminals 1 and 2) must ALWAYS be connected for the 8850 to operate.

Technical Notes

Display Condition

Possible Causes

Suggested Solutions

"- - - -"

Display is out of range. This may be a normal 
condition if your process operates at or near 
the limits of the electrode range.

•  Check sensor for correct range.
•  Check Decimal setting in OPTIONS menu.
•  Check Calibrate menu settings for incompatible 

SOURCE and RANGE values.

"Value must be

less than 3"

The PPM factor must be a value from
0.00 to 3.00.

Set a PPM value less than 3.

Value must be greater than 0

Custom cell constant cannot be set to 0

Set cell constant to a value greater than 0

"Value must be

400 or less"

Pulse Rate cannot be greater than 400.

Set a pulse rate value less than 400.

Display erratic,

shows "0"  or "------"

intermittently

Aux Power not connected.

Connect Aux. Power.
See sections 3.1 and 3.2 for wiring information.

"Too much Error

Check Sensor"

•  Defective or dirty/coated sensor.
• Wiring error.
•  Fluid conductivity too low.
•  Defective temp element.

•  Clean or replace sensor.
•  Check and correct wiring.
• Replace sensor.

Содержание Signet ProcessPro 8850-2

Страница 1: ...nder ISO 9001 and ISO 14001 CAUTION Remove power to unit before wiring input and output connections Follow instructions carefully to avoid personal injury Contents 1 Installation 2 Specifications 3 El...

Страница 2: ...val Procedure 1 Press the orange terminal lever downward with a small screwdriver to open terminal jaws 2 When fully open remove wire from terminal Wiring Tips Do not route sensor cable in conduit con...

Страница 3: ...Loop AUX Power AUX Power Transmitter 2 Terminals PLC Terminals Loop Input 4 20 mA Transmitter Terminals System Pwr Loop System Pwr Loop AUX Power AUX Power Power Supply 12 24 VDC Power Supply 12 24 VD...

Страница 4: ...keys The selections will scroll in a continuous loop Changing the VIEW display does not interrupt system operations No key code is necessary to change display selection Output settings cannot be edite...

Страница 5: ...2 The Key Code is keys in sequence After entering the Key Code the display will show the first item in the selected menu Step 3 Scroll menu with or arrow keys Step 4 Press key to select menu item to b...

Страница 6: ...ut Loop Range S 0 0000 100 00 Select the minimum and maximum values for the 4 to 20 mA current loop output Be sure to modify this setting if you change the Cond Units Relay1 Source Cond Select Conduct...

Страница 7: ...able display and output response Temp Display C Select temperature units C or F Temperature Comp 2 00 Set the percent change in Conductivity caused by a 1 C change in temperature May be from 0 00 to 1...

Страница 8: ...t value If the instrument does not display the correct value Temperature 0 5 C Conductivity 2 of reading service is required A Accuracy Verification with Precision Resistors Electronic Calibration 1 S...

Страница 9: ...ayed temperature and conductivity values in the spaces provided Displayed temperature T1 _______ C Displayed conductivity C1 _______ S 3 Cool the sample solution close to the minimum process temperatu...

Страница 10: ...condition if your process operates at or near the limits of the electrode range Check sensor for correct range Check Decimal setting in OPTIONS menu Check Calibrate menu settings for incompatible SOUR...

Страница 11: ...11 8850 2 Conductivity Resistivity Transmitter Notes...

Страница 12: ...085 CR Sanitary 0 01 SS 1 to 1 1 2 in 3 2819 S1C 159 000 087 CR Sanitary 0 01 SS 1 to 1 1 2 in Certified 3 2819 S2 159 000 086 CR Sanitary 0 01 SS 2 in 3 2819 S2C 159 000 088 CR Sanitary 0 01 SS 2 in...

Отзывы: