7
8850-2 Conductivity/Resistivity Transmitter
Options Menu
Display
(Factory settings shown)
Description
Contrast:
3 >
Adjust the LCD contrast for best viewing. A setting of 1 is lowest contrast, 5 is highest. Select
lower contrast if the display is in warmer ambient surroundings.
Cond Decimal:
****.* >
Set the decimal to the best resolution for your application. The display will automatically scale
up to this restriction. Select *****., ****.*, ***.** **.*** or *.****.
Averaging
Off >
OFF provides the quickest output response to changes in temperature. Longer averaging
period produces more stable display and output response.
Temp Display
°C >
Select temperature units: ºC or ºF.
Temperature Comp %:
2.00 >
Set the percent change in Conductivity caused by a 1 ºC change in temperature. May be from
0.00 to 10.00.
Loop Adjust
4.00 mA
>
Adjust the minimum and maximum current output. Use this setting to match the system output
to any external device. The display value represents the precise current output.
Adjustment limits:
• 3.80 mA < 4.00 mA > 5.00 mA
• 19.00 mA < 20.00 mA > 21.00 mA
Loop Adjust
20.00 mA
>
Test Loop:
>
Press
▲
or
▼
keys to manually order any output current value from 3.6 mA to 21.00 mA to test
current loop output.
Test Relay 1:
>
Press
▲
or
▼
keys to manually toggle the state of Relay 1.
Test Relay 2:
>
Press
▲
or
▼
keys to manually toggle the state of Relay 2.