background image

8850-2 Conductivity/Resistivity Transmitter 

Calibration Procedure

This instrument has been electronically calibrated at the factory.
• Procedure A veri

fi

 es the accuracy and linearity of the instrument by simulating temperature and conductivity values with precision 

(±0.1%) 

fi

 xed resistors.

•  Procedure B is a wet calibration. This procedure uses the sensor input and NIST traceable test solutions. When done correctly, this 

procedure offers the most accurate system calibration.

Iso. Gnd 

(BLACK) 

 

Temp. IN 

(WHITE) 

 

Signal IN 

(RED) 

13

12

11

Simulation resistor

TC resistor 

B) Wet Calibration with NIST Traceable Solutions:

When using NIST traceable standards, review the temperature information provided with the test solution. Prevent 
contamination of the test solution. The sensor must be at the temperature speci

fi

 ed on the test solution label.

•  Remove the sensor from the system. Rinse the sensor in a small amount of test solution.
•  Place the sensor into the test solution.
•  Place a reference thermometer into the same solution.
• Allow suf

fi

 cient time for the temperature to stabilize.

•  Set Temp: Adjust the temperature value based on the reference thermometer. (see Editing Procedure.)
•  Set Cond:  Adjust the conductivity value based on the test solution value. (see Editing Procedure.)
•  Verify the linearity by placing the sensor into a second test solution of a different value.
•  If the instrument does not display the correct value (Temperature ± 0.5 ºC, Conductivity ± 2% of reading), service is required.

A) Accuracy Veri

fi

 cation with Precision Resistors (Electronic Calibration):

1.  Simulate the Temperature

The temperature input to the 8850 is a PT-1000 thermistor, where 1000 Ohms (

Ω

) is equal to 0 ºC and a change of 3.84 

Ω

 

equals a 1 ºC change. (1000 

Ω

 = 0 ºC, 1003.84 

Ω

 = 1.0 ºC, 1007.68 

Ω

 = 2.0 ºC. ... 1096 

Ω

 = 25 ºC)

•  Connect a resistor (1000 

Ω

 to 1096 

Ω

 ) between "Temp IN" and "Iso. Gnd" terminals.

•  Set Temp; Adjust the temperature to exact value based on the measured resistance. (see Editing Procedure, Calibrate menu).
•  To verify the temperature linearity, connect a second resistor value to the terminals.
•  If the instrument does not display the correct value, service is required. ( ±0.5 ºC)

2.  Simulate the Conductivity

You may calculate the exact Resistance needed to simulate a speci

fi

 c conductivity value , or you may calculate the exact 

Conductivity based on a resistor value:

(*1 

μ

S = 1 X 10

-6

 Siemens or 0.000001 Siemens)

•  Connect the conductivity resistor between the "Signal IN " and "Iso Gnd" terminals.
•  Set Cond: Adjust the conductivity value based on the resistor value. (see Editing Procedure 

and Calibrate menu).

•  Verify the linearity of the instrument by connecting a second resistor of a different value.
•  If the instrument does not display the correct value (± 2% of reading), service is required.

Resistance =

Cell constant

e.g.

0.1 Cell 

= 5,000 

Ω

 or 5 K

Ω

conductivity (Siemens*)

0.000020 (Siemens*)

Conductivity =

Sensor cell

e.g.

0.1 Cell

= 0.000001 Siemens* or 1

μ

S/cm

Simulation resistance (

Ω

)

100,000 (

Ω

)

Содержание Signet ProcessPro 8850-2

Страница 1: ...nder ISO 9001 and ISO 14001 CAUTION Remove power to unit before wiring input and output connections Follow instructions carefully to avoid personal injury Contents 1 Installation 2 Specifications 3 El...

Страница 2: ...val Procedure 1 Press the orange terminal lever downward with a small screwdriver to open terminal jaws 2 When fully open remove wire from terminal Wiring Tips Do not route sensor cable in conduit con...

Страница 3: ...Loop AUX Power AUX Power Transmitter 2 Terminals PLC Terminals Loop Input 4 20 mA Transmitter Terminals System Pwr Loop System Pwr Loop AUX Power AUX Power Power Supply 12 24 VDC Power Supply 12 24 VD...

Страница 4: ...keys The selections will scroll in a continuous loop Changing the VIEW display does not interrupt system operations No key code is necessary to change display selection Output settings cannot be edite...

Страница 5: ...2 The Key Code is keys in sequence After entering the Key Code the display will show the first item in the selected menu Step 3 Scroll menu with or arrow keys Step 4 Press key to select menu item to b...

Страница 6: ...ut Loop Range S 0 0000 100 00 Select the minimum and maximum values for the 4 to 20 mA current loop output Be sure to modify this setting if you change the Cond Units Relay1 Source Cond Select Conduct...

Страница 7: ...able display and output response Temp Display C Select temperature units C or F Temperature Comp 2 00 Set the percent change in Conductivity caused by a 1 C change in temperature May be from 0 00 to 1...

Страница 8: ...t value If the instrument does not display the correct value Temperature 0 5 C Conductivity 2 of reading service is required A Accuracy Verification with Precision Resistors Electronic Calibration 1 S...

Страница 9: ...ayed temperature and conductivity values in the spaces provided Displayed temperature T1 _______ C Displayed conductivity C1 _______ S 3 Cool the sample solution close to the minimum process temperatu...

Страница 10: ...condition if your process operates at or near the limits of the electrode range Check sensor for correct range Check Decimal setting in OPTIONS menu Check Calibrate menu settings for incompatible SOUR...

Страница 11: ...11 8850 2 Conductivity Resistivity Transmitter Notes...

Страница 12: ...085 CR Sanitary 0 01 SS 1 to 1 1 2 in 3 2819 S1C 159 000 087 CR Sanitary 0 01 SS 1 to 1 1 2 in Certified 3 2819 S2 159 000 086 CR Sanitary 0 01 SS 2 in 3 2819 S2C 159 000 088 CR Sanitary 0 01 SS 2 in...

Отзывы: