background image

7

2750 DryLoc pH/ORP Sensor Electronics

•  When the 2750 is powered with 12 to 24 VDC, the 4 to 20 mA loop output is automatically selected.
•  Remove approximately 10 mm (0.4 in.) of insulation and tin each conductor before inserting into connectors. 
•  To add the EasyCal function to the 2750-1, 2750-3 and 2750-4, use the 8050-2 or 8052-2 junction box.

Current Loop with no junction box

Current loop with Junction box

4 to 20 mA Loop Wiring

All 2750 pH/ORP Sensor 
Electronics are factory-calibrated for 
maximum out-of-the-box accuracy. 
Periodic calibration is required to 
compensate for electrode aging. 

The optional EasyCal feature allows 
calibration to be performed local to 
the sensor.

EasyCal Procedure:

The 

fi

 rst step (Reset) is recommended each time an electrode is replaced, but is NOT necessary upon initial installation or periodic 

calibration.
In fact, for periodic calibration it is best for the electrode/sensor assembly to remain intact to minimize the possibility of moisture or other 
contamination entering the electrical interconnection area. The electrode/sensor connection must remain dry and clean at all times.
1.  Reset the 2750 pH/ORP Sensor Electronics to factory calibration: With no electrode connected to the Sensor, press and hold SW1 

until the LED (D1) comes on steady then goes off again (approx. 10 seconds). When the LED goes off, release SW1; reset is complete.

2.  Connect an electrode to the 2750 pH/ORP Sensor Electronics.

3. If 

digital 

(S

3

L) output is being used, place SW2 in the “Cal” position. If 4 to 20 mA output is being used, SW2 position is of no 

consequence.

4.  Place the electrode/sensor assembly into a calibration solution as follows: (If the electrode is “healthy”, then the 2750 will automatically 

recognize the solution. The order in which the solutions are used during the calibration procedure is of no consequence). 

•  For pH calibration, use any two of these international standards: pH 4.0, 7.0 or 10.0 buffer solutions.

(Signet part number 3-0700.390 contains one capsule of each value)

•  To produce standards for ORP calibration, mix the chemical Quinhydrone into pH 7.0 and 4.0 buffers to saturation (

 g per 50ml). 

•  Regardless of the size of the container used for calibration, one inch of solution is adequate to completely submerge the tip of the 

electrode.

• 

Allow at least 30 seconds for the electrode response to stabilize before calibration.

5.  Press and hold SW1 for approximately 8 to 10 seconds.  During this time, the LED (D1) will come on steady then go back off. 

Release SW1 (If the LED blinks several times rapidly, the calibration was not successful. See the troubleshooting section.)

6.  Remove the electrode/sensor assembly from the 

fi

 rst calibration solution, rinse the electrode with clean water, and place it in a 

second solution.  

Allow at least 30 seconds for the electrode response to stabilize before calibration.

7.  Press and hold SW1 for approximately 8 to 10 seconds.  During this time, the LED will come on steady then go back off. 

Release SW1. (If the LED blinks several times rapidly, the calibration was not successful. See the troubleshooting section.)

8. 

For 

digital 

(S

3

L) systems ONLY: Return SW2 to the RUN position.

Calibration is complete. Return the system to service.

Calibration

SW1

 EasyCal 

button

SW2

  RUN/CALIBRATE switch for (S

3

L) system

D1

   

EasyCal Indicator (green LED)

SW1

D1

+

S

-

SW2

Run

Cal

8052-2

8050-2

SW1

D1

SW2

Run

Cal

1

2

3

4

+

S

-

2750-2

Shield

Red

No Connection

Black

White

Power Supply

12 to 24 VDC

+

-

PLC or Recorder

Loop Input 

4 to 20 mA

-

+

Optional Earth ground

+

S

-

1

2

3

4

Black

Red

Power Supply

12 to 24 VDC

+

-

PLC or Recorder

Loop Input 

4 to 20 mA

-

+

2750-1

1

2

3

4

+

S

-

2750-2

SW1

D1

SW2

Run

Cal

1

2

3

4

+

S

-

8050-2, 8052-2

SW1

D1

+

S

-

SW2

Run

Cal

8050-1, 8052-1

+

S

-

Optional Earth ground

Required Equipment:

•  

2750-2 with built-in EasyCal 
or 

•  

Any 2750 with 8050-2 or 8052-2 junction box with EasyCal

•  Standard pH buffers (pH 4, 7 or 10)

Содержание Signet 2750-1

Страница 1: ...4X Junction Boxes are integral parts of the in line versions 2750 1 2750 2 and 2750 7 and are available as accessories for the submersible versions 2750 3 and 2750 4 The optional EasyCal feature allow...

Страница 2: ...00 or 3K Balco temperature sensor versions for pH Operational Range 2750 0 00 to 14 00 pH 2000 mV ORP 0 C to 85 C 32 F to 185 F Response Time includes electrode response pH 6 s for 95 of change ORP ap...

Страница 3: ...A Signet installation fitting 12 7 mm to 101 6 mm 1 2 in to 4 in or a threaded tee is required to secure the electrode in the pipe The Signet Measurement and Instrumentation catalog offers a complete...

Страница 4: ...nts or their vapor may lead to failure of the retaining nut ejection of the sensor and loss of the process fluid with possibly serious consequences such as damage to equipment and serious personal inj...

Страница 5: ...othing when working with chemicals in pressurized pipe Locked The molded alignment marks are enhanced for clarity 2750 In Line Assembly 1 Insert electrode into base of the 2750 3 or 2750 4 and turn un...

Страница 6: ...00 users If this direct wiring is used set the CALIBRATE menu to perform Calibration at INSTRUMENT Connect the terminals as shown to any digital S3 L terminals 8900 and 9900 users If using EasyCal set...

Страница 7: ...consequence For pH calibration use any two of these international standards pH 4 0 7 0 or 10 0 buffer solutions Signet part number 3 0700 390 contains one capsule of each value To produce standards f...

Страница 8: ...tor kit PG 13 5 1 connector 3822 7004 159 001 581 pH 4 01 buffer solution 1 pint 473 ml bottle 3822 7007 159 001 582 pH 7 00 buffer solution 1 pint 473 ml bottle 3822 7010 159 001 583 pH 10 00 buffer...

Отзывы: