2750 DryLoc pH/ORP Sensor Electronics

Caution / Warning / Danger

Indicates a potential hazard. Failure to follow all warnings 
may lead to equipment damage, injury, or death

Personal Protective Equipment (PPE)

Always utilize the most appropriate PPE during 
installation and service of Signet products.

Pressurized System Warning

Sensor may be under pressure, take caution to vent 
system prior to installation or removal. Failure to do so 
may result in equipment damage and/or serious injury.

Hand Tighten Only

Overtightening may permanently damage product threads 
and lead to failure of the retaining nut.

Do Not Use Tools

Use of tool(s) may damage product beyond repair and 
potentially void product warranty.

Note / Technical Notes

Highlights additional information or detailed procedure.

  Safety Information

Refer to your local Georg Fischer Sales of

fi

 ce for the most 

current warranty statement.

All warranty and non-warranty repairs being returned must 
include a fully completed Service Form and goods must be 
returned to your local GF Sales of

fi

 ce or distributor. 

Product returned without a Service Form may not be 
warranty replaced or repaired.

Signet products with limited shelf-life (e.g. pH, ORP, chlorine 
electrodes, calibration solutions; e.g. pH buffers, turbidity 
standards or other solutions) are warranted out of box but not 
warranted against any damage, due to process or application 
failures (e.g. high temperature, chemical poisoning, dry-out) or 
mishandling (e.g. broken glass, damaged membrane, freezing 
and/or extreme temperatures).

Thank you for purchasing the Signet line of Georg Fischer 
measurement products.
If you would like to register your product(s), you can now register 
online in one of the following ways:
•  

Visit our website 

www.gfsignet.com. 

Under 

Service and 

Support

 click on 

Product Registration Form

•  If this is a pdf manual (digital copy), 

click here

General

Compatible Electrodes .........  Signet DryLoc Electrodes 

(PT1000 or 3K Balco temperature 
sensor versions for pH)

Operational Range 2750 ......  0.00 to 14.00 pH, ±2000 mV ORP, 

0 °C to 85 °C (32 °F to 185 °F)

Response Time (includes electrode response):
  pH ..................................... <6 s for 95% of change
 ORP 

.................................. application 

dependent

  pH Temp T90 ...................  200 s (2724 and 2734)

132 s (2726 and 2736)
438 s (2756-WT)

Materials

In-line: 
  2750-1, 2750-2, 2750-7.....  PBT (thermal plastic polyester)
Submersible: 
 2750-3, 

2750-4 

.................. CPVC

Cable for 2750-3, 2750-4, 2750-7: 
  •

 

5 m (15 ft) 3-conductor shielded, 22 AWG

  •

 

May be extended up to 305 m (1000 ft) with current output

  •  May be extended up to 305 m (1000 ft) with (S

3

L) output

Shipping Weight

2750-1, 2750-2 ...................... 0.75 kg (1.75 lb)
2750-3, 2750-4, 2750-7 ........ 0.64 kg (1.4 lb)

Environmental

Ambient Temp. ..................... 0 °C to 85 °C (32 °F to 185 °F)
Storage Temp. .....................  -20 °C to 85 °C (-4 °F to 185 °F)
Relative Humidity ................. 95% max, non-condensing
Immunity ............................... EN50082-2
Emissions ............................. EN50081-1

Enclosure Rating

In-line: 
 2750-1, 

2750-2, 2750-7 ....  NEMA 4X/IP65 

(with electrode connected)

Submersible:
 2750-3, 

2750-4 

.................  

NEMA 6P/IP68 
(with electrode and watertight 
extension pipe connected)

Electrical

Input Impedance .................. >10

11 

Ω

Input response time ............. 500 ms
Temperature drift..................  ±0.002 pH per ºC,  

±0.1 mV ORP per ºC

Input resolution ..................... 0.02 pH, 1 mV ORP, 0.3 ºC

Specifi cations

Chemical Compatibility

Warranty Information

Product Registration

1.  Depressurize and vent system prior to installation or removal.
2. Con

fi

 rm chemical compatibility before use.

3.  Do not exceed maximum temperature/pressure speci

fi

 cations.

4.  Wear safety goggles or faceshield during installation/service.
5.  Do not alter product construction.
6.  When using chemicals or solvents care should be taken 

and appropriate eye, face, hand, body, and/or respiratory 
protection should be used.

The retaining nuts of pH and ORP sensors are not designed 
for prolonged contact with aggressive substances. 

Strong acids, caustic substances and solvents 
or their vapor may lead to failure of the retaining 
nut, ejection of the sensor and loss of the process 

fl

 uid with possibly serious consequences, such as 

damage to equipment and serious personal injury. 

Retaining nuts that may have been in contact with 
such substances e.g. due to leakage or spilling, 
must be replaced.

Содержание Signet 2750-1

Страница 1: ...4X Junction Boxes are integral parts of the in line versions 2750 1 2750 2 and 2750 7 and are available as accessories for the submersible versions 2750 3 and 2750 4 The optional EasyCal feature allow...

Страница 2: ...00 or 3K Balco temperature sensor versions for pH Operational Range 2750 0 00 to 14 00 pH 2000 mV ORP 0 C to 85 C 32 F to 185 F Response Time includes electrode response pH 6 s for 95 of change ORP ap...

Страница 3: ...A Signet installation fitting 12 7 mm to 101 6 mm 1 2 in to 4 in or a threaded tee is required to secure the electrode in the pipe The Signet Measurement and Instrumentation catalog offers a complete...

Страница 4: ...nts or their vapor may lead to failure of the retaining nut ejection of the sensor and loss of the process fluid with possibly serious consequences such as damage to equipment and serious personal inj...

Страница 5: ...othing when working with chemicals in pressurized pipe Locked The molded alignment marks are enhanced for clarity 2750 In Line Assembly 1 Insert electrode into base of the 2750 3 or 2750 4 and turn un...

Страница 6: ...00 users If this direct wiring is used set the CALIBRATE menu to perform Calibration at INSTRUMENT Connect the terminals as shown to any digital S3 L terminals 8900 and 9900 users If using EasyCal set...

Страница 7: ...consequence For pH calibration use any two of these international standards pH 4 0 7 0 or 10 0 buffer solutions Signet part number 3 0700 390 contains one capsule of each value To produce standards f...

Страница 8: ...tor kit PG 13 5 1 connector 3822 7004 159 001 581 pH 4 01 buffer solution 1 pint 473 ml bottle 3822 7007 159 001 582 pH 7 00 buffer solution 1 pint 473 ml bottle 3822 7010 159 001 583 pH 10 00 buffer...

Отзывы: