2750 DryLoc pH/ORP Sensor Electronics

Mounting Position

Mounting Angle using GF Signet Fittings.

*

Avoid locations with air pockets and/or sediment (90° and 270°).

Wet-Tap electrodes, 2724, 2734, and 2774 series electrodes

•   Vertical (0°) position optimum.
•    DO NOT install within 30º of horizontal (Contact factory 

for horizontal or inverted installation requirements).

2764 series electrodes MUST be mounted upright.

When mounting in a standard threaded 

fi

 tting, the electrode must 

be mounted horizontal to 60 degrees below horizontal position only:

2. 

Thread the in-line retaining nut onto 

fi

 tting to secure the electrode into place.

Chemical Compatibility Warning

• 

The retaining nuts of pH and ORP sensors are not designed for prolonged contact 
with aggressive substances. 

• 

Strong acids, caustic substances and solvents or their vapor may lead to failure of 
the retaining nut, ejection of the sensor and loss of the process 

fl

 uid with possibly 

serious consequences, such as damage to equipment and serious personal injury. 

• 

Retaining nuts that may have been in contact with such substances e.g. due to 
leakage or spilling, must be replaced.

1.

  Insert electrode into Signet installation 

fi

 tting.  

Seat the electrode tabs into the alignment notches in the 

fi

 tting.

NOTE:

This procedure applies to systems using 2724-2726 and 2734-2736 electrodes. 
If the 3719 pH/ORP Wet-Tap is used, refer to the 3719 manual for instructions.

2750 In-Line Assembly

30°

30°

Vertical ± 60°

YES

YES

NO

NO

NO

NO

NO

Lubricate O-rings with a 
non-petroleum based, 
viscous lubricant 
(grease) compatible 
with the system.

HAND-TIGHTEN THE THREADED NUT ONTO THE INSTALLATION FITTING.
DO NOT USE TOOLS! DO NOT USE THREAD SEALANT OR LUBRICANTS 
ON THE FITTING THREADS OR THE SENSOR CAP.

Retaining nut

270°

90°

180°

300°

240°

YES

YES

NO

NO

NO

NO

OK

OK

OK

OK

OK

OK

OK

OK

*

*

Содержание Signet 2750-1

Страница 1: ...4X Junction Boxes are integral parts of the in line versions 2750 1 2750 2 and 2750 7 and are available as accessories for the submersible versions 2750 3 and 2750 4 The optional EasyCal feature allow...

Страница 2: ...00 or 3K Balco temperature sensor versions for pH Operational Range 2750 0 00 to 14 00 pH 2000 mV ORP 0 C to 85 C 32 F to 185 F Response Time includes electrode response pH 6 s for 95 of change ORP ap...

Страница 3: ...A Signet installation fitting 12 7 mm to 101 6 mm 1 2 in to 4 in or a threaded tee is required to secure the electrode in the pipe The Signet Measurement and Instrumentation catalog offers a complete...

Страница 4: ...nts or their vapor may lead to failure of the retaining nut ejection of the sensor and loss of the process fluid with possibly serious consequences such as damage to equipment and serious personal inj...

Страница 5: ...othing when working with chemicals in pressurized pipe Locked The molded alignment marks are enhanced for clarity 2750 In Line Assembly 1 Insert electrode into base of the 2750 3 or 2750 4 and turn un...

Страница 6: ...00 users If this direct wiring is used set the CALIBRATE menu to perform Calibration at INSTRUMENT Connect the terminals as shown to any digital S3 L terminals 8900 and 9900 users If using EasyCal set...

Страница 7: ...consequence For pH calibration use any two of these international standards pH 4 0 7 0 or 10 0 buffer solutions Signet part number 3 0700 390 contains one capsule of each value To produce standards f...

Страница 8: ...tor kit PG 13 5 1 connector 3822 7004 159 001 581 pH 4 01 buffer solution 1 pint 473 ml bottle 3822 7007 159 001 582 pH 7 00 buffer solution 1 pint 473 ml bottle 3822 7010 159 001 583 pH 10 00 buffer...

Отзывы: