28
CS100/CS300 BATTERY OPERATION
Operation from DC-to-AC Inverter
The CS100 can be powered from a DC-to-AC inverter if the DC source and the inverter meet
the specifications defined below. The DC source and inverter should be checked for proper
operation by qualified maintenance personnel prior to emergency use. The message Battery
in Use will not be displayed during proper AC inverter operation.
Specifications for DC Source and Inverter for Use with the CS100
Voltage Output:
100-120/220-240 VAC ± 10%
Frequency:
50 Hz ±2 Hz, 60 Hz ±2 Hz
Overshoots:
Does not continuously generate overshoots greater than
375. Volts peak with widths greater than 10 mSec when
powering the system.
Waveform:
Sine wave or modified sine wave
Output Capability:
Minimum of 500 watts continuous power; 1000 watts
surge power
Safety Compliance:
Meets or exceeds safety standards per IEC 60601-1.
CS100 Battery
Operation
Содержание CS100