13 CS100/CS300 BATTERY OPERATION
Informational Messages
Battery Status
Associated Help Screen
Probable Cause
Corrective Actions
The IABP internal battery requires
maintenance.
1. Contact Getinge Service.
2. Continue operation on AC power.
3. If battery operation is necessary, battery run time may be
reduced. Monitor the system for a Low Battery alarm.
NOTE: If a Low Battery alarm occurs, immediately connect the
system to AC power.
CAUSE AND RESPONSE:
Operation
Mode
Trigger
Source
Detailed Cause
System Response
Reset
All
All
The Battery Test Due Date
or Battery Replacement
Date predate the current
system date at startup
or the internal battery
has a total accumulated
discharge time in excess of
100 total discharge cycles.
None
At startup the message
is cleared when the
START key is pressed.
During operation the
message is removed
after 30 seconds.
BATTERY MAINTENANCE REQUIRED