18
CS100/CS300 BATTERY OPERATION
CS300 Battery Operation
CS300 Battery
Operation
Battery in Use
• Verify that the “Battery in Use” Informational Message and the Battery Indicator are displayed
when in battery operation.
• Battery charging is not active when the battery is in use.
• When the battery has approximately 30 minutes of operating time remaining, the following occurs:
• The Medium Priority Alarm tone is activated.
• The “Low Battery” Medium Priority Alarm message is displayed on the screen continuously.
• The Battery Indicator is displayed as empty and it starts flashing.
• The condensate removal module will not operate (to conserve remaining power).
A reduction in run time will occur over a battery’s life due to age, storage temperature and discharge cycles.
Batteries which are continually subjected to complete discharge cycles without the recommended
immediate recharging, can incur permanent damage.