19 CS100/CS300 BATTERY OPERATION
CS300 Operation from DC-to-AC Inverter
CS300 Battery
Operation
The IABP can be powered from a DC-to-AC inverter if the DC source and the inverter meet the specifications
defined below. The DC source and inverter should be checked for proper operation by qualified maintenance
personnel prior to emergency use. The message Battery in Use will not be displayed during proper AC
inverter operation.
Specifications for DC Source and Inverter for Use with the IABP
Voltage Output:
100-120/220-240 VAC ± l 0%
Frequency:
50 Hz ±2 Hz, 60 Hz ±2 Hz
Overshoots:
Does not continuously generate overshoots greater than 375
Volts peak with widths greater than 10 mSec when powering the system
Waveform:
Sine wave or modified sine wave
Output Capability:
Minimum of 500 watts continuous power; 1000 watts surge power
Safety Compliance:
Meets or exceeds safety standards per IEC 60601-1