23 CS100/CS300 BATTERY OPERATION
CS100 Portable Operation
The following is recommended by Datascope during portable operation:
1. Battery fully charged prior to starting transport.
2. Altitude changes are compensated for automatically in the Autofill mode or manually in the Manual Fill mode.
3. The system must be properly secured when used in an ambulance, helicopter, or fixed wing aircraft.
The CS100 is available in two versions with identical clinical functionality. A console version with a hospital cart
or the CS100 Universal Transport System (UTS) is specifically designed for air and vehicular transport. The UTS
version mechanically attaches to a docking station for high strength mounting.
WARNING: The user should continually rely on visual alarm messages during high noise transport
situations. The “Flash Alarms” option should be turned ON to improve the visibility of alarm messages.
This option can be set in the User Preferences Menu (located in the Display Preferences submenu.)