NRG 16-19, NRG 16-27, NRG 16-28
Service pressure
PN 40, max. 32 barg at 238°C
Connection
NRG 16-19, NRG 16-27: screwed
3
/
8
" to A ISO 228
NRG 16-28: screwed M 24 x 1.5
Materials
NRG 16-19
Screw-in body: 1.4301
Spacer disk: 1.4571
Electrode rod: 1.4571
Insulating disk: Gylon
®
NRG 16-27, NRG 16-28
Screw-in body: 1.4571
Electrode rod: 1.4571
Insulation of electrode rod: PEEK
Sensitivity of response
> 1 μS/cm at 25°C
Electrode voltage
12 V
Electrical connection
NRG 16-19: Connecting line made from PTFE, 2 m long, 2 x 1.5 mm
2
NRG 16-27, NRG 16-28: M 12 sensor connector, 5 poles, A coded
Protection
NRG 16-27, NRG 16-28: IP 65 to DIN EN 60529
NRG 16-19: IP 52 to DIN EN 60529
Ambient temperature at terminal box
Max. 80°C
Weight
Approx. 0.3 kg
Technical Data
Gylon
®
is a registered trademark of Garlock GmbH, Neuss
Corrosion resistance
When used for its intended purpose the safe functioning of the electrode will not be impaired by
corrosion.
Sizing
The body is not designed for pulsating loads. Dimensional allowances and anti-corrosive additives are
in accordance with the latest state-of-the-art.