Instructions for Use
Ergostik
Serial no.: xx|8|201|yyy and 2201xxxxx
Version: 7 | Release date: 02. August 2021
Page 73
9.2.3
Removing the Silicone Mask / Adapter
1.
Remove the adapter.
(In reserve order as described in chap.
7.4.1 “Assembly Face Mask“)
2.
Wipe off visible contamination. with a
soft cloth.
3.
Insert the adapter and silicone mask
into a disinfectant solution.
4.
Wash the adapter and silicone mask
thoroughly in distilled water so that
no residues of contamination and
disinfectants remain
5.
Air-dry adapter and silicone mask
until no moisture is visible.