![GERATHERM 979119 Скачать руководство пользователя страница 52](http://html1.mh-extra.com/html/geratherm/979119/979119_instructions-for-use-manual_2216238052.webp)
Instructions for Use
Ergostik
Serial no.: xx|8|201|yyy and 2201xxxxx
Page 52
Version: 7 | Release date: 02. August 2021
Before the Ergostik can be used to perform measurements
again, all components must be properly reconnected, the USB
connection to the computer must be established, and the
Ergostik must be connected to the power supply and calibrated.
Calibration also fulfils the regulatory requirements of MPBetreibV
§7 for functional testing after maintenance work.
Please proceed as described in the respective chapters.
6.2.1
Assembly Calibration Tube
Possible physical injury.
Reason: Measurement error due to leakage caused by
incorrect assembly of components. Therefore:
•
Carefully observe the assembly instructions!
Attention
Functional disorders possible.
Reason: Components and e.g. plug connections can be
damaged. Therefore:
•
All connections must be made carefully and without too
much force!