![GERATHERM 979119 Скачать руководство пользователя страница 35](http://html1.mh-extra.com/html/geratherm/979119/979119_instructions-for-use-manual_2216238035.webp)
Instructions for Use
Ergostik
Serial no.: xx|8|201|yyy and 2201xxxxx
Version: 7 | Release date: 02. August 2021
Page 35
3.2
The Technical State of Ergostik and System Construction
Possible danger to life.
Reason: Electric shock. Cross-contamination.
Misdiagnosis caused by measurement error.
Therefore:
•
Do not overstress Ergostik! Use with care!
•
Do not modify or use product Ergostik or total
system contrary to the respective manufacturer's
specification!
•
Only use Ergostik within its expected service life,
determined by the manufacturer!
•
Under no circumstances use or connect any
devices, systems, equipment and other products
that are not part of the total system!
•
Never obstruct the access to the mains plug or
On / Off switch of the equipment cart!
Disconnection of the power supply must be easily
accessible!
•
Only use accessories and consumables which are
authorised by the manufacturer as well as
authorised replacement components, add-ons
and auxiliary materials.
Check that the components are in a functional
and safe condition!
•
Ensure that in the patient environment – a
distance of 1.5 meters to the patient – there are
no accessible electrical parts (interfaces, plugs,
etc.) that are not isolated from the mains with an
isolation voltage of 4 kV!
•
Ensure that the users or third party persons do
not touch the patient and any conductive
connections or parts of the device that are
located outside the patient environment at the
same time!