![GERATHERM 979119 Скачать руководство пользователя страница 68](http://html1.mh-extra.com/html/geratherm/979119/979119_instructions-for-use-manual_2216238068.webp)
Instructions for Use
Ergostik
Serial no.: xx|8|201|yyy and 2201xxxxx
Page 68
Version: 7 | Release date: 02. August 2021
Attention
Ergostik could be damaged.
Reason: Penetrating liquid into electrical components.
Therefore:
•
Disconnect the Ergostik from the power supply before
cleaning and disinfecting!
•
(When switching off via the On / Off switch of the
equipment cart), shut down the PC completely before
doing so!
•
Wipe off remaining moisture with a dry cloth!
Ergostik could fail.
Reason: Damaged connections. Therefore:
•
To disconnect electrical connections, always pull the
plug and never the cable!
9.1
Single Use
For instructions on handling noseclips, read:
•
Chap. 7.7 “Exchange of Disposable Products /
Disinfection”
9.2
Disinfection
All parts of the Ergostik can be cleaned of dirt with a soft cloth
using a cleaning solution / weak soapy water.
All parts of Ergostik which come into contact, or could come into
contact, with the patient must be treated with a surface
disinfection. Otherwise, these can be wiped with a soft cloth
using a weak soap solution.