Attention
Independent of this, the user / operator must carry out regular
checks during day-to-day operation, see also the following
chap. 8 “Servicing / Maintenance by the User”!
8.2
Servicing / Maintenance by the User / Operator
In order to ensure a flawless operation of Ergostik over its whole
service life, regular servicing and repairs, if applicable, are
required.
Intervall
Servicing work
Before each
examination
Checking the cable and tube connections
Once per day
Visual checks of the device and its components for
damage and replacing them if necessary
(see chap. 8.2.1 “Checking / Exchanging of Tubes and
Cables”)
After about 50 to
100 examinations
Replacing the permapure tube and filter
After 12 months
Replacing the oxygen cell and the exercise tube set