November 2004
1-34
Programming Site Data
Frequency Outputs
The frequency output issues a continuous signal with a frequency
proportional to the selected measurement. While following the
programming instructions, refer to Figure A-4 in Appendix A.
Complete the following steps to set up the totalizer outputs of an
option card installed in
SLOT 1
:
Accessing the Frequency Output
1.
Open the
SITE EDIT MENU
from the
Edit Functions
menu.
2.
Select
Global
.
3.
Select
Input/Output
.
4.
Select
Options
.
5.
Select
Slot 1
.
6.
Use the pull-down menu to select the desired input and press
Next
Item/Enter
.
7.
Do one of the following:
•
If you selected
Off
, IDM returns to the
Input/Output
window.
Go to
Procedure Options on
the following page.
•
If you selected
Frequency
, and are using a
•
1-Channel meter, proceed to Step 2 in the next section.
•
2-Channel meter, proceed to Step 1 in the next section.
Note:
Output letters (A, B, etc.) correspond to the output wiring
terminal numbers from top to bottom. For example, Output A
corresponds to pins 1, 2 and 3; Output B corresponds to pins
4, 5 and 6; etc.
Setting Up the Frequency
1.
Use the pull-down menu to select the desired channel option and
press
Next Item/Enter
.
2.
Use the pull-down menu to select the desired measurement
parameter and press
Next Item/Enter
. See Table 1-12 on the next
page for a description of the available options. Refer to page 3-2 in
the
Service Manual
for a description of diagnostic parameters.