Programming Site Data
1-31
November 2004
Selecting the Alarm Type (cont.)
4.
Do one of the following:
•
If you are using a 1-Channel meter and selected
•
high or low, proceed to Step 7.
•
fault, proceed to Step 9.
•
If you are using a 2-Channel meter, proceed to the next step.
5.
Use the pull-down menu to select the desired channel option and
press
Next Item/Enter
.
6.
Do one of the following:
•
high or low, proceed to Step 7.
•
fault, proceed to Step 9.
7.
Use the pull-down menu to select the desired measurement
parameter and press
Next Item/Enter
. See Table 1-10 below for a
description of the available options.
The
DIAG*
option listed above represents all of the individual
diagnostic parameters that appear at the previous prompt. See
Chapter 3,
Diagnostics
, of the
Service Manual
for a complete
description of these options.
Note:
The measurement units that appear in these prompts are those
selected in the
Global-System
window earlier in this section.
8.
Enter a value for the trigger point of the alarm and press
Next
Item/Enter
. Go to
Procedure Options
on the following page.
9.
Use the pull-down menu to select the type of error (flow, non-flow
or both) that will trigger the fault alarm and press
Next Item/
Enter
.
Table 1-10: Output Measurement Options
Options
Flow Velocity
Volumetric Flow
Forward Totalized Volume Flow
Reverse Totalized Volume Flow
Total Flow Measurement Time
Mass Flow
Forward Totalized Mass Flow
Reverse Totalized Mass Flow
Diagnostic Parameters*