Programming Site Data
1-9
November 2004
Pipe Data (cont.)
4.
Enter the axial length of the ultrasonic signal, then use the pull-
down menu to select the units. Press
Next Item/Enter
.
5.
Use the pull-down menu to select the fluid type and press
Next
Item/Enter
. Then do one of the following:
•
If
OTHER
was selected - proceed to Step 6.
•
If
STEAM
was selected - proceed to Step 7.
6.
Enter the speed of sound (in feet per second) in the gas to be
measured and press
Next Item/Enter
.
7.
Enter a value for the flow calibration factor and press
Next Item/
Enter
. The default value is 1.00, but values between 0.50 and 2.0
may be entered.
Procedure Options
After entering the
Calibration Factor
, IDM returns to the
Channel
PROGRAM
. Do one of the following:
•
To continue entering “quick startup” data, press
EXIT PAGE
once
and proceed to Step 2 in
Entering Global System Data
on
.
•
To continue regular programming, refer to Appendix A,
, to navigate to the desired menu.
•
To leave the
User Program
, press the
EXIT PAGE
key twice.