November 2004
1-16
Programming Site Data
Set Transducer Signal
Settings (cont.)
Table 1-4: Transducer Signal Settings
Transducer
Signal
Parameters
Range
Default
Value
Miscellaneous Information
Signal Low
Limit
-20 to 100
20
The E1:LOW SIGNAL error message appears when the
signal strength falls below the programmed SIGNAL LOW
LIMIT value. See Chapter 2,
Error Codes
, in the
Service
Manual
for a discussion of error codes.
Correlation
Peak Limit
0 to 500
100
The E4: SIGNAL QUALITY error message appears when the
signal quality falls below the programmed COR. PEAK
LIMIT value. See Chapter 2,
Error Codes
, in the
Service
Manual
for a discussion of error codes
Soundspeed
+- Limit
1 to 50%
20%
The E2:SOUNDSPEED error message appears when the
calculated fluid sound speed differs from the fluid sound
speed entered in the Channelx-System menu by more
than the programmed SOUN- LIMIT value.
See Chapter 2,
Error Codes
, in the
Service Manual
for a
discussion of error codes.
Velocity Low
Limit
-500 to
500 ft/sec
(-150 to
150 m/sec)
-75.0 ft/sec
(-23 m/sec)
The E3: VELOCITY RANGE error messages appears when
the calculated fluid velocity is less than the programmed
VELOCITY LOW LIMIT value.
See Chapter 2,
Error Codes
, in the
Service Manual
for a
discussion of error codes.
Velocity High
Limit
-500 to 500
ft/sec (-150
to 150 m/
sec)
75.0 ft/sec
(23 m/sec)
The E3: VELOCITY RANGE error messages appears when
the calculated fluid velocity exceeds the programmed
VELOCITY HIGH LIMIT value.
See Chapter 2,
Error Codes
, in the
Service Manual
for a
discussion of error codes.
Acceleration
Limit
0 to 250ft/s
(0 to 76m/s)
50 ft/s
(15 m/s)
The E6: CYCLE SKIP error message appears when the
calculated fluid velocity changes by more than the
programmed ACCELERATION LIMIT value from one reading
to the next. See Chapter 2,
Error Codes
, in the
Service
Manual
for a discussion of error codes.
Amplitude
Discriminator
Low
0 to 100
14
The amplitude discriminator measures the transducer
signal received by the XGS868. The default value for this
parameter is 14, and values from 0 to 100 are acceptable.
The E5: AMPLITUDE error message appears when the
amplitude discriminator falls below the programmed AMP.
DISCRIM LOW value. See Chapter 2,
Error Codes
, in the
Service Manual
for a discussion of error codes.
Amplitude
Discriminator
High
0 to 100
34
The amplitude discriminator measures the transducer
signal received by the Model XGS868. The default value for
this parameter is 34. Values from 0 to 100 are acceptable.
The E5: AMPLITUDE error message appears when the
amplitude discriminator exceeds the programmed AMP.
DISCRIM HIGH value. See Chapter 2,
Error Codes
, in the
Service Manual
for a discussion of error codes.
Содержание XGS868
Страница 6: ...Chapter 1 ...
Страница 48: ...Chapter 2 ...
Страница 55: ...Chapter 3 ...
Страница 66: ...Appendix A ...
Страница 74: ...Appendix B ...
Страница 75: ...Data Records Available Option Cards B 1 Option Cards Installed B 2 Setup Data B 3 ...
Страница 81: ...Appendix C ...
Страница 104: ...Appendix D ...