January 2008
6-2
XDP Specifications
Functional Specifications
Alarms (terminal blocks J2, J3):
Dual general-purpose contacts,
Contact rating: 2 A, 28 V, SPDT
XDP Fault (J3):
fail-safe type
XDP Normal (J3):
fail-safe type
XDP Low (J3):
fail-safe or non-fail-safe mode,
trips if H2 <85% (setpoint is programmable)
XDP Low-Low (J3):
fail-safe or non-fail-safe mode,
trips if H2 <80% (setpoint is programmable)
Analog Input (terminal block J6):
4–20 mA, 28 V
Analog Output (terminal block J8):
4–20 mA, 28 V
Communications (terminal block J4):
Standard RS232/RS485 Serial Port
Display:
4-line, backlit liquid crystal display (LCD)
Keypad:
6-key, infrared
Measurement Ranges (for sample application):
•
70–100% hydrogen in air
•
0–100% hydrogen in carbon dioxide
•
0–100% air in carbon dioxide
Environmental
Specifications
The design, manufacture, and certification of the XDP have been
conducted in a manner that allows its operation in Zone 1,
Gas Group IIC, flameproof protection method in an ambient
operating temperature of –10° to 60°C (14° to 140°F).
This unit complies with EMC Directive 89/336/EEC and the
73/23/EEC Low Voltage Directive (Installation Category II, Pollution
Degree 2). It is in compliance with ATEX directive 94/9/EC Annex
II, per the document at the back of this manual.
Transmitter
Specifications
For complete transmitter specifications, refer to the documentation
provided with the transmitter.
Содержание XDP-H2
Страница 3: ...Chapter 1 ...
Страница 4: ...General Information Introduction 1 1 Basic Features 1 1 System Overview 1 2 A Typical Application 1 3 ...
Страница 9: ...Chapter 2 ...
Страница 10: ...Installation Introduction 2 1 Choosing an Installation Site 2 2 Mounting the XDP System 2 3 Wiring the XDP 2 4 ...
Страница 20: ...Chapter 3 ...
Страница 28: ...Chapter 4 ...
Страница 55: ...Chapter 5 ...
Страница 76: ...Chapter 6 ...
Страница 80: ...Appendix A ...
Страница 81: ...CE Mark Compliance Introduction A 1 EMC Compliance A 1 LVD Compliance A 1 ...
Страница 83: ...Appendix B ...
Страница 84: ...Additional Wiring Connections Introduction B 1 Connecting the System Outputs B 1 XDP Alarm Relays B 2 ...
Страница 87: ...Appendix C ...