January 2008
5-16
Advanced Programming
Configuring the 4-20 mA
Analog Input (cont.)
Now, the XDP analog output is driven to exactly 20 mA.
To continue programming the setup menu, proceed to the appropriate
section for instructions. Otherwise, press
[ESC]
four times to return to
live measurement mode.
Entering Factory
Calibration Data
To enter the factory calibration data, access the setup menu as
described on page 5-4 and complete the following steps:
At the next prompt, select the range to be calibrated (H2/AIR is used
as an example in this manual).
At the next prompt, enter the number of data points available for
calibrating the chosen gas curve.
4-20mA-I..20mA-Trim
Press the [
S
] and [
T
] keys
until the meter reads exactly
20 mA. When done, press
[ENTER]
.
xx.xxx
s
xx
t
4-20mA-Input
Press
[ESC]
to return to the
setup submenu.
Test
Trim
Range
Cal
Setup
Use the [
S
] and [
T
] keys to
select
[Factory Calibration]
and
press
[ENTER]
.
Serial Number
4-20mA-Input
Factory Calibration
Gas Curve:
Use the [
S
] and [
T
] keys to
select
[H2/AIR]
and press
[ENTER]
.
H2/AIR
H2/CO2
AIR/CO2
Gas-Curve: H2/AIR
Use [
W
] and [
X
] to position the
cursor under the desired char-
acter. Then, use [
Œ
] and [
´
] to
increment the value. Press
[ENTER]
when done.
Enter # of Points
xx
s
Содержание XDP-H2
Страница 3: ...Chapter 1 ...
Страница 4: ...General Information Introduction 1 1 Basic Features 1 1 System Overview 1 2 A Typical Application 1 3 ...
Страница 9: ...Chapter 2 ...
Страница 10: ...Installation Introduction 2 1 Choosing an Installation Site 2 2 Mounting the XDP System 2 3 Wiring the XDP 2 4 ...
Страница 20: ...Chapter 3 ...
Страница 28: ...Chapter 4 ...
Страница 55: ...Chapter 5 ...
Страница 76: ...Chapter 6 ...
Страница 80: ...Appendix A ...
Страница 81: ...CE Mark Compliance Introduction A 1 EMC Compliance A 1 LVD Compliance A 1 ...
Страница 83: ...Appendix B ...
Страница 84: ...Additional Wiring Connections Introduction B 1 Connecting the System Outputs B 1 XDP Alarm Relays B 2 ...
Страница 87: ...Appendix C ...