![GE XDP-H2 Скачать руководство пользователя страница 39](http://html.mh-extra.com/html/ge/xdp-h2/xdp-h2_user-manual_128493039.webp)
January 2008
4-10
Setup and Calibration
Setting the Zero Setpoint
Enter the system analog output menu, as described on page 4-8, and
proceed as follows:
Note:
The
Test
,
Zero Trim
, and
Span Trim
options are also available
in the above list box.
At the next prompt, enter the desired zero setpoint (the percentage of
target gas that corresponds to the lower end of the specified system
analog output range).
Either press
[ESC]
until you return the
User Menu
or proceed to the
appropriate section to continue setting up the system analog output.
User..Aout
Use the [
S
] and [
T
] keys to
select
[Zero Setpoint]
, and press
[ENTER]
.
Aout Type
Zero Setpoint
Span Setpoint
User Aout Zero-Set
Use [
W
] and [
X
] to position the
cursor under the desired char-
acter. Then, use [
S
] and [
T
] to
increment the value. Press
[ENTER]
when done.
Zero Set
xxxx.xxxxx
s
User..Aout
You are now back at the sys-
tem analog output setup menu.
Aout Type
Zero Setpoint
Span Setpoint
Содержание XDP-H2
Страница 3: ...Chapter 1 ...
Страница 4: ...General Information Introduction 1 1 Basic Features 1 1 System Overview 1 2 A Typical Application 1 3 ...
Страница 9: ...Chapter 2 ...
Страница 10: ...Installation Introduction 2 1 Choosing an Installation Site 2 2 Mounting the XDP System 2 3 Wiring the XDP 2 4 ...
Страница 20: ...Chapter 3 ...
Страница 28: ...Chapter 4 ...
Страница 55: ...Chapter 5 ...
Страница 76: ...Chapter 6 ...
Страница 80: ...Appendix A ...
Страница 81: ...CE Mark Compliance Introduction A 1 EMC Compliance A 1 LVD Compliance A 1 ...
Страница 83: ...Appendix B ...
Страница 84: ...Additional Wiring Connections Introduction B 1 Connecting the System Outputs B 1 XDP Alarm Relays B 2 ...
Страница 87: ...Appendix C ...