January 2008
5-6
Advanced Programming
The Range Submenu
(cont.)
The Test Submenu
At the next prompt, enter the mA value to be output as a test signal.
The above two prompts repeat so that more than one test value may
be tried. When the testing is complete, press
[ESC]
from the first
prompt to leave the
Test
submenu.
4-2..4-20_Out_Range
Press
[ESC]
to leave the
[4-20
Out_Range]
submenu.
H2/AIR
H2/CO2
AIR/CO2
4-20mA-Out
Use the [
S
] and [
T
] keys to
select
[Test]
and press
[ENTER]
.
4-20_Out_Range
Test
Trim
4-2..Enter-mA-Value
Use [
W
] and [
X
] to position the
cursor under the desired char-
acter. Then, use [
S
] and [
T
] to
increment the value and press
[ENTER]
when done.
Enter-mA-Value
xx.xx
s
4-20mA-Out En..Test
Measure the XDP analog out-
put test signal and compare it
to the displayed value. When
done, press
[ENTER]
.
Enter-mA-Value
xx.xx
s
Содержание XDP-H2
Страница 3: ...Chapter 1 ...
Страница 4: ...General Information Introduction 1 1 Basic Features 1 1 System Overview 1 2 A Typical Application 1 3 ...
Страница 9: ...Chapter 2 ...
Страница 10: ...Installation Introduction 2 1 Choosing an Installation Site 2 2 Mounting the XDP System 2 3 Wiring the XDP 2 4 ...
Страница 20: ...Chapter 3 ...
Страница 28: ...Chapter 4 ...
Страница 55: ...Chapter 5 ...
Страница 76: ...Chapter 6 ...
Страница 80: ...Appendix A ...
Страница 81: ...CE Mark Compliance Introduction A 1 EMC Compliance A 1 LVD Compliance A 1 ...
Страница 83: ...Appendix B ...
Страница 84: ...Additional Wiring Connections Introduction B 1 Connecting the System Outputs B 1 XDP Alarm Relays B 2 ...
Страница 87: ...Appendix C ...