Technical Data / Information
7.13 Color
Doppler
C-Mode:
Color flow imaging is possible with Curved Array, Linear Array and Phased Array probes.
Display Modes:
2D/CFM (Single, Dual, Quad); 2D+2D/C
Simultaneous triplex mode 2D/CFM/D
Color coding steps:
8192 color steps
Depth range:
axial: 0 to B-scan range
lateral: 0 to B-scan-range
Zero line shift:
17 steps (independent from spectral Doppler)
Inversion of color direction:
yes
Wall Motion Filter:
7 steps
Smoothing Filter:
12 steps rising time
12 steps falling time
Gain control :
30dB
Density (color line density):
10 steps
Ensemble (color shots per line): 7 to 31
Pulse repetition frequency:
100 Hz to 20.5 kHz
CFM Map:
8 different color codes for each probe
Frequency range:
1 to 15 MHz depending on the probe,
adjustable in 3 steps (low, mid, high)
Balance:
from 25 to 225 in 41 steps
Max. meas. velocity:
5.5 m/sec.
Min. meas. velocity:
0.3 cm/sec.
Display Modes:
V-T (ve turbulence)
V
(velocity)
V-P (ve power)
T
(turbulence)
P-T (power + turbulence)
Scale:
kHz,
cm/s,
m/s
Automatic moving tissue suppression: yes
Voluson
®
730 - Instruction Manual
7-10
105838 Rev. 3
Содержание Voluson 730
Страница 4: ...This page intentionally left blank Voluson 730 Instruction Manual i 2 105838 Rev 3 ...
Страница 9: ...General 1 General 1 2 Voluson 730 Instruction Manual 105838 Rev 3 1 1 ...
Страница 48: ...Description of the System Voluson 730 Instruction Manual 3 18 105838 Rev 3 This page intentionally left blank ...
Страница 62: ...Electronic User Manual EUM Voluson 730 Instruction Manual 5 8 105838 Rev 3 This page intentionally left blank ...
Страница 66: ...Connections 6 2 1 Main Module Voluson 730 Instruction Manual 6 4 105838 Rev 3 ...
Страница 79: ...Connections 6 2 11 ECG preamplifier MAN 6 Connection Voluson 730 Instruction Manual 105838 Rev 3 6 17 ...
Страница 82: ...Connections This page intentionally left blank Voluson 730 Instruction Manual 6 20 105838 Rev 3 ...