Connections
6.3
Important Notes: Connecting Auxiliary Equipment
CAUTION !
The leakage current of the entire system including any / all auxiliary
equipment must not exceed the limit values as per EN60 601-1-1:1990 (IEC
60601-1-1) resp. other valid national or international standards.
All equipment must comply with UL, CSA and IEC requirements.
CAUTION !
Auxiliary equipment must only be connected to the main console with the
special mains outlet provided for the electrical safety of the system.
CAUTION !
Auxiliary equipment with direct mains connection requires galvanic
separation of the signal and / or control leads.
CAUTION:
The outlets ST1 – ST5 may be set to 230V or 115V
Modification of voltage setting only by an authorized service person!
ATTENTION ! With the color video monitor connected to its isolated mains outlet, the
remaining load capacity for auxiliary equipment is max. 350 VA.
WARNING !
Only accessories explicitly recognized by the system manufacturer
GE Medical Systems Kretztechnik GmbH & Co OHG may be used in
connection with the system.
Voluson
®
730 - Instruction Manual
105838 Rev. 3
6-19
Содержание Voluson 730
Страница 4: ...This page intentionally left blank Voluson 730 Instruction Manual i 2 105838 Rev 3 ...
Страница 9: ...General 1 General 1 2 Voluson 730 Instruction Manual 105838 Rev 3 1 1 ...
Страница 48: ...Description of the System Voluson 730 Instruction Manual 3 18 105838 Rev 3 This page intentionally left blank ...
Страница 62: ...Electronic User Manual EUM Voluson 730 Instruction Manual 5 8 105838 Rev 3 This page intentionally left blank ...
Страница 66: ...Connections 6 2 1 Main Module Voluson 730 Instruction Manual 6 4 105838 Rev 3 ...
Страница 79: ...Connections 6 2 11 ECG preamplifier MAN 6 Connection Voluson 730 Instruction Manual 105838 Rev 3 6 17 ...
Страница 82: ...Connections This page intentionally left blank Voluson 730 Instruction Manual 6 20 105838 Rev 3 ...