Chapter 8 - Replacement Procedures
8-121
P R E L I M I N A R Y
8-6-3-5
Battery Bracket Installation Procedure
NOTE:
When performing the following installation procedure steps, it may be helpful to review the appropriate
movie clips previously referenced in the
8-6-3-4 "Battery Bracket Removal Procedure" on page 8-119
1.) Working with the battery bracket upside down on the table surface (and close to the system),
reconnect the battery cable (attached to the battery bracket) to the BEP (see
).
2.) Carefully lift the battery bracket back and turning it over, swing into position on the cage assembly
(see
).
3.) Slide the bracket in place and make sure it is seated correctly.
4.) Return and fasten the four screws (previously removed) to secure the battery bracket to the cage
assembly, as indicated in
5.) Return the battery, as described in the
8-6-2-5 "Battery Installation Procedure" on page 8-118
6.) Carefully secure the cables with new cable tie wraps.
7) Return the cabinet cage assembly to the system, reconnect the cables (previously disconnected),
close the cabinet cage door and fasten securely with the two latches.
Refer to the
8-6-1-4 "Cabinet Cage Assembly Closing and Installation Procedure" on page 8-114
8) Return the front cover, as described in the
8-2-6-4 "Front Cover Installation Procedure" on page 8-
.
9) Return the left and right side covers, as described in the
8-2-3-4 "Left Side Cover Installation
and See “Right Side Cover Installation Procedure” on page 5..
10.)Turn ON power to the system.
11.)Proceed to perform the following functionality tests:
-
10-6-4 "Grounding Continuity" on page 10-23
-
Содержание Vivid S5 N
Страница 2: ......
Страница 4: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ii P R E L I M I N A R Y ...
Страница 5: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iii P R E L I M I N A R Y ...
Страница 6: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iv P R E L I M I N A R Y ...
Страница 7: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL v P R E L I M I N A R Y ...
Страница 8: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vi P R E L I M I N A R Y ...
Страница 9: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vii P R E L I M I N A R Y ...
Страница 10: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL viii P R E L I M I N A R Y ...
Страница 11: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ix P R E L I M I N A R Y ...
Страница 12: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL x P R E L I M I N A R Y ...
Страница 13: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xi P R E L I M I N A R Y ...
Страница 14: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xii P R E L I M I N A R Y ...
Страница 833: ......
Страница 834: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 NO 3191 HORTEN NORWAY www gehealthcare com ...