Section 8-6 - Lower Section Components Replacement Procedures
P R E L I M I N A R Y
8-6-2 Battery Replacement Procedure
8-6-2-5
Battery Installation Procedure
1) Holding the cabinet cage assembly steady with one hand, carefully slide the replacement battery
sideways into the battery compartment.
2.) Make sure the battery is properly inserted all the way, before proceeding to the next step (refer to
3) Return the metal securing bracket (previously removed) fasten in position with the attached screw
).
4) Return the cabinet cage assembly to the system, reconnect the cables (previously disconnected),
close the cabinet cage door and fasten securely with the two latches.
Refer to the
8-6-1-4 "Cabinet Cage Assembly Closing and Installation Procedure" on page 8-114
5) Return the Vivid S6 N front cover, as described in the
8-2-6-4 "Front Cover Installation Procedure"
6) Return the Vivid S6 N left and right side covers, as described in the
and
8-2-2-4 "Right Side Cover Installation Procedure" on page 8-5
.
7.) Turn ON power to the Vivid S6 N ultrasound scanner.
8.) Proceed to perform the following functionality tests:
-
10-6-4 "Grounding Continuity" on page 10-23
-
10-6-5 "Chassis Current Leakage Test" on page 10-24
-
7-2-4-2-1 "Calibration" on page 7-12
-
7-2-4-1 "Full System Test" on page 7-11
WARNING
WARNING
The Vivid S6 N battery has a safety device.
DO NOT ATTEMPT TO DIS-ASSEMBLE OR ALTER THE BATTERY
!
Always observe the precautions listed in
1-3-5 "Vivid S5 N / Vivid S6 N Battery Safety" on page 1-
CAUTION
use only batteries approved by GE HEALTHCARE as suitable for use with the Vivid S6 N
ultrasound scanner
Содержание Vivid S5 N
Страница 2: ......
Страница 4: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ii P R E L I M I N A R Y ...
Страница 5: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iii P R E L I M I N A R Y ...
Страница 6: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iv P R E L I M I N A R Y ...
Страница 7: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL v P R E L I M I N A R Y ...
Страница 8: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vi P R E L I M I N A R Y ...
Страница 9: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vii P R E L I M I N A R Y ...
Страница 10: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL viii P R E L I M I N A R Y ...
Страница 11: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ix P R E L I M I N A R Y ...
Страница 12: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL x P R E L I M I N A R Y ...
Страница 13: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xi P R E L I M I N A R Y ...
Страница 14: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xii P R E L I M I N A R Y ...
Страница 833: ......
Страница 834: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 NO 3191 HORTEN NORWAY www gehealthcare com ...