Chapter 10 - Care and Maintenance
10-27
10-6-7
Isolated Patient Lead (Source) Leakage – Lead-to-Lead
NOTE:
Refer to the procedure in the IEC 60601-1.
NOTE:
Electrical leakage testing may be accomplished with any calibrated Electrical Safety Analyzer
tool compliant with AAMI/ESI 1993 or IEC 60601 or AS/NZS 3551.
10-6-8
Isolated Patient Lead (Sink) Leakage - Isolation Test
NOTE:
Refer to the procedure in the IEC 60601-1.
NOTE:
Electrical leakage testing may be accomplished with any calibrated Electrical Safety Analyzer
tool compliant with AAMI/ESI 1993 or IEC 60601 or AS/NZS 3551.
NOTE:
It is not necessary to test each lead individually, or power condition combinations, as required
in previous tests.
CAUTION
Line voltage is applied to the ECG leads during this test. To avoid the risk of possible electric
shock, the system being tested must not be touched by anyone (patient or operator), while the
ISO TEST switch is depressed.
Содержание Vivid q N
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type...
Страница 697: ......
Страница 698: ...2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com...