Chapter 4 - General Procedures and Functional Checks
4-29
4-3-4
Back End Processor Tests
4-3-4-1
Magneto Optical Drive (MOD) Test
Table 4-18 MOD Test - Recommended
Step
Task
Expected Result(s)
1
Restart the system performing a full shutdown and
bootup, as described in
.
During the boot-up process, verify that the messages
SCSI Adaptor Installation 2271149-2INS
and
MO
are
displayed.
2
Insert a new MO Media Sony EDM-2300B (2.3G) into
the MOD.
3
Press
Ctrl+F11
(or
Alt+D
)
on the keyboard and select
<Back End>, <Media Driver> and <MO>, as appropriate.
The UtilMediaApp
dialog box is displayed.
4
Select the
IsMediaAcces
button.
The message
Media in drive G is accessible
is
displayed.
5
Select the
EjectMO
button.
The message
Media is outside
is displayed. Check
that the media has actually been ejected.
6
Select the
LoadMO
button.
The message
Media is inside
is displayed. Check that
the media is actually inside.
7
Select the
Format
button.
The message
Media in drive G was formatted
successfully
is displayed.
8
Select the
GetVolInfo
button.
The message
Media in drive G has following
properties: Media Label = MY_LABEL,
Media File-System Name = FAT
is displayed.
9
Select the
Get
BTPsector
button.
The message
The Bytes Per Sector value of media
in drive G is 512
is displayed.
10
Select the
SetVolLabel1
button.
The message
Media Label set to TEST_LABEL1
is
displayed.
11
Select the
GetVolInfo
button.
The message
Media in drive G has following
properties: Media Label = TEST_LABEL,
Media File-System Name = FAT
is displayed.
12
Select the
Exit
button and close the Diagnostic
Application.
13
Press the
Archive
button on the keyboard.
The
Patient List
window is displayed.
14
Press the
Utilities
button on the soft menu.
The
General Status
window is displayed.
15
Press the
Removable Media Utility
button.
The
Utility
window is displayed, showing the message:
Free space on Media: 1096 MB
.
16
Type the new label (for example, “Side1”), then select
the
Format
button.
Confirm “YES” to the warning message.
The message
Current Media Name is:SIDE1. Media
was formatted successfully
is displayed.
MO Label: SIDE1, MO STATUS: MOUNTED
17
Select the
Eject
button.
The message
Media Ejected
is displayed.
MO STATUS: Ejected.
18
Select the
OK
button to exit the utility.
Содержание Vivid q N
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type...
Страница 697: ......
Страница 698: ...2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com...