Chapter 10 - Care and Maintenance
10-17
10-5-4-2
General Probe Cleaning Instructions
10-5-4-2-1
Cleaning and Sterilizing Probes Used Externally
10-5-4-2-2
Cleaning and Sterilizing Probes Used Internally
WARNING
WARNING
Neurological procedures must NOT be performed on a patient with Creutzfield-Jacob
disease because probes that have been used for such examinations cannot be
adequately sterilized.
WARNING
WARNING
Neurological and intraoperative procedures must be performed with the use of legally
marketed, sterile, pyrogen-free probe sheaths. Probes used during neurological
surgery must NOT be sterilized with liquid chemical sterilants because of the possibility
of neurotoxic residue remaining on the probe.
Table 10-11 Cleaning & Sterilizing Procedure for Probes Used Externally
Step
Description
1
To protect yourself from blood-borne diseases, wear approved, non-allergic disposable gloves.
2
Do not immerse the connector or its strain relief in solution. Use ultrasound probe cleaning towels.
Wash the probe and probe cable in warm soap and water solution (below 80
°
F). Use only mild
non-abrasive soap.
3
Follow the manufacturer’s cold sterilization instructions to disinfect the probe, including concentration
and soaking time; 10 minutes is a typical disinfecting period.
4
Follow step 3 for a much longer duration to sterilize the probe; 10 hours is a typical cold chemical
sterilization duration.
5
When complete, thoroughly rinse the washed lens in clear water. Apply sterile distilled water for at
least 60 seconds. Air dry.
6
Check the lens for cracking and the cable insulation for separation.
7
Store the probe in its carrying case.
Table 10-12 Cleaning & Sterilizing Procedure for Probes (TEE) Used Internally
Step
Description
1
To protect yourself from blood-borne diseases, wear approved, non-allergic disposable gloves.
2
Do not immerse the connector or its strain relief in solution. Use ultrasound probe cleaning towels.
Wash the probe and probe cable in warm soap and water solution (below 80
°
F). Use only mild
non-abrasive soap.
3
Follow the manufacturer’s cold sterilization instructions to disinfect the probe, including concentration
and soaking time; 10 minutes is a typical disinfecting period.
4
Follow step 3 for a much longer duration to sterilize the probe; 10 hours is a typical cold chemical
sterilization duration.
5
When complete, thoroughly rinse the washed lens in clear water. Apply sterile distilled water for at
least 60 seconds. Air dry.
6
Check the lens for cracking and the cable insulation for separation.
7
Store the probe in its carrying case.
Содержание Vivid q N
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type...
Страница 697: ......
Страница 698: ...2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com...