Section 3-10 - Connectivity Setup
3-10-8
eVue Dataflow Configuration
3-10-8-1
Introduction
It is possible to configure the Vivid q N portable ultrasound scanner to allow the user to send images or
loops to a remote PC, in almost real time. When pressing Store, the loop will save not only to the local
archive (in raw-data format), but will also be transmitted to a designated PC (“e-Vue PC”) on the network
(in compact MPEG format), and will appear at the review PC within a very short time.
This feature proves very useful in the OR scenario, where consultation with another doctor in the
hospital
in real-time
is required.
3-10-8-2
Preparations
Before configuring the Vivid q N for eVue dataflow, it is necessary to decide upon a specific PC in the
hospital as being the e-Vue PC station. The service technician then configures the
Vivid q N to send images to that specific e-Vue PC, as described below.
3-10-8-3
Configuring the Vivid q N for eVue Dataflow
1.) Verify which PC in the hospital/institute has been designated as the e-Vue station.
2.) Make sure the Vivid q N is connected to the Network via cable or wireless.
3.) On the Vivid q N scanner, press the
Config
button.
The System Configuration window opens.
4.) Click
Connectivity
.
The Connectivity dialog box opens.
5.) Select the
Dataflow
Tab, as shown in
.
Содержание Vivid q N
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type...
Страница 697: ......
Страница 698: ...2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com...