3900/3900P User’s Manual
2-4
Averaging mode
Switches 5 and 6 set the averaging mode. The averaging mode selects the time
period of data used to calculate a weighted average SpO
2
value to be displayed by
the oximeter.
Switches
SpO
2
and Pulse Rate Averaging
TruTrak
+
5 6
5 6
Long (12 seconds)
Yes (enabled)
5 6
Medium (6 seconds)
No
5 6
Short (3 seconds)
No
P
I
r
pulsatile value display
Switch 7 sets the display of the PI
r
pulsatile value.
Switch
Display P
I
r
value
7
Yes
7
No
EMI line frequency
Switch 8 sets the EMI line frequency. To optimize EMI (electromagnetic
interference) immunity, make sure switch 8 is in the correct position for the AC
power line frequency in use.
Switch
EMI Line Frequency
8
60 Hz
8
50 Hz