3900/3900P User’s Manual
2-14
NOTE: If an alarm occurs while you are entering label characters, the waveform
screen appears with the alarm message. After you have attended to the alarm
condition, return to the LABELS screen. The characters you had entered prior to
the alarm are still there but they are not yet saved in memory as the label for the
current data record. You do not have to reenter them, but you do need to
complete the label information and then press
to save the label to memory.
CLEAR
Select CLEAR to access the clear patient data screen options:
Option
Action
CLEAR LABELS
Clears label text from unlocked lines on the LABELS
screen.
CLEAR TREND
Clears all SpO
2
and PI
r
trend data in memory, including
the labels in memory associated with that data.
CLEAR BOTH
Clears label text from unlocked lines on the LABELS
screen
and
clears all stored trend data, including labels
in memory associated with that data.
After choosing an option, you can confirm your choice by selecting YES or NO.
•
If you select YES, a high beep tone sounds. If you select NO, a low beep tone
sounds. A message indicating that the labels, trends, or both have or have not
been cleared (depending on your selection) is displayed for three seconds.
•
If you do not select YES or NO within 10 seconds, a low beep tone sounds and
the previous screen is displayed.
NOTE: Text on the LABELS screen that is not erased will be used for the next
data record.
Locking label lines
Line 1 remains in memory until you change it (or clear it) and press
. This
line has 13 character fields and it cannot be locked.
Lines 2, 3, and 4 can each be locked to retain their information until you unlock
the lines and change the text or use CLEAR to remove it. These lines have 12
character fields each. The first character field on each line is reserved for a lock or
unlock symbol. With the cursor on the symbol, use the –/+ function keys to switch
between the symbols and thereby lock or unlock that line. When a line is locked,
it is not erased from the screen when you select CLEAR.
Содержание TruTrak+ 3900
Страница 1: ...GE Healthcare 3900 3900P Pulse Oximeter TruTrak User s Manual ...
Страница 2: ......
Страница 3: ...GE Healthcare 3900 3900P Pulse Oximeter TruTrak User s Manual 6050 0006 406 March 2005 ...
Страница 26: ...3900 3900P User s Manual 1 16 ...
Страница 80: ...3900 3900P User s Manual 5 18 ...
Страница 96: ...A 4 3900 3900P User s Manual ...
Страница 102: ...3900 3900P User s Manual B 6 ...
Страница 106: ...Index ...