Chapter 8. Servicing the PT878
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
161
8.4.1 Entering the Material and Soundspeed (cont.)
If you have selected a preprogrammed material, you have completed
entering data in this window. Pressing the [
T
] key returns the program to
the Material tab. But if you selected “Other,” you can also enter a specific
soundspeed.
3.
To enter the soundspeed:
a.
Press
[ENTER]
to open the text box.
b.
Use the numeric keys to enter the desired soundspeed (from the
brochure
Soundspeeds and Pipe Size Data,
914-004).
c.
Press
[ENTER]
to confirm your entry.
To leave the Material window:
•
Press the [
W
] or [
X
] arrow key to move to another tab to take
readings or perform a velocity calibration.
•
Press
[F2]
(Cancel) to return to Operate Mode without confirming the
soundspeed value.
•
Press
[F3]
(OK) to confirm the new value. The PT878 returns to
Operate Mode.