3.11.3 Entering a Calibration Factor
The Calibration Factor is used to calibrate or adjust the readings of the
PT878 to another flow reference. To enter a Calibration Factor, return to the
Reynolds Correction tab at the top of the Correction Factors option. Press
the [
X
] arrow key to reach the Calibration Factor window. The window
appears similar to Figure 36 below. To step through each parameter, press
the [
T
] key.
Figure 36: The Calibration Factor Window
1.
The first prompt asks if you wish to enable the calibration factor.
a.
Use the [
W
] and [
X
] keys to scroll to the appropriate radio button.
b.
Press
[ENTER]
to confirm your selection.
2.
The next prompt asks if you want a single K factor or a table of K
factors.
a.
Use the [
W
] and [
X
] keys to scroll to the appropriate radio button.
b.
Press
[ENTER]
to confirm your selection.
Reynolds Correctio
Inputs
Correction Factors
Calibration Factor
Edit Table
Calibration Factor
On
Off
Meter K-Factor
Data Source
K Factor
Single
Table
10
Edit Table
Cancel
OK