Chapter 8. Servicing the PT878
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
167
Dual-Point Calibration (cont.)
IMPORTANT:
Be sure to hold the transducer steady until the GE wait
cursor disappears.
10.
Use the [
S
] key to return to the Zero tab, and then scroll to the Display
tab (page 162) to confirm the block length within ±0.002. If the
measurement is not within this limit, recalibrate the transducer offsets.
Once you have completed calibration, the transducer should remain
calibrated until the meter memory is cleared or a dramatic shift in ambient
conditions occurs.
To leave the Zero window:
•
Press the [
W
] or [
X
] arrow key to move to another tab to take
readings or perform a velocity calibration.
•
Press
[F2]
(Cancel) to return to Operate Mode without confirming the
soundspeed value.
•
Press
[F3]
(OK) to confirm the new value. The PT878 returns to
Operate Mode.