Chapter 3. Programming Site Data
66
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
3.11.2 Entering Reynolds Correction (cont.)
1.
The first prompt asks if you want to enable the Reynolds Correction
factor, a number based on the Kinematic Viscosity and flow rate of the
fluid.
Note:
If you are using clamp-on transducers, you should enable Reynolds
Correction.
a.
Use the [
W
] and [
X
] keys to scroll to the appropriate radio button.
b.
Press
[ENTER]
to confirm your selection.
Note:
If you disable Reynolds Correction, you will not be able to enter any
other values in this window.
2.
If you enable Reynolds Correction, you will also need to enter the
Kinematic Viscosity of the fluid (available in the brochure
Soundspeeds
and Pipe Size Data
, 914-004). The prompt asks for a kinematic
viscosity value.
a.
Press
[ENTER]
to open the text box.
b.
Use the numeric keys to enter the desired value (from
Soundspeeds
and Pipe Size Data
).
c.
Press
[ENTER]
to confirm your selection.
You have completed entering data for a single Kinematic Viscosity factor.
Proceed to
Entering a Calibration Factor
on page 67.
•
To confirm the entries and return to Operate Mode, press
[F3]
(OK). The
PT878 returns to Operate Mode.
•
To leave the window without confirming the entries, press
[F2]
(Cancel)
or the
[ESC]
key.