Chapter 9. Diagnostics and Troubleshooting
206
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
9.2 Displaying Diagnostic Parameters
As part of its measurement menu, the PT878 offers a list of diagnostic
parameters to aid in troubleshooting in the event of flowcell, transducer, or
electrical problems. You can select any diagnostic parameter for display as a
measurement as discussed in Chapter
5, Displaying and Configuring Data,
on page 95. Table 8 below shows all the available diagnostic parameters
and ranges.
Table 8: Diagnostic Parameters
Diagnostic
Parameter
Displays
Good
Bad
DT
Displays Delta T, or the difference between
the upstream and downstream transit
times.
N/A
Continuous
large fluctu-
ations of 1
ms or more.
Amplitude
Up
Displays the value for the amplitude dis-
criminator of the upstream transducer.
20-28
<20 or >28
Amplitude
Down
Displays the value for the amplitude dis-
criminator of the downstream transducer.
20-28
fluctuations
< 20 or > 28
fluctuations
T Up
Displays the upstream transit time of the
ultrasonic signal in micro seconds.
N/A
N/A
T Down
Displays the downstream transit time of
the ultrasonic signal in micro seconds.
N/A
N/A
Gain/Up
Displays upstream gain in dB.
N/A
N/A
Gain/Dn
Displays downstream gain in dB.
N/A
N/A
Soundspeed
(m/s or f/s)
Displays the measured soundspeed of the
fluid.
Check the soundspeed of
fluid in
Soundspeeds and
Pipe Size Data.
P#up
Displays signal peaks for the upstream
transducer.
100-900
<100, >900
P#dn
Displays signal peaks for the downstream
transducer.
100-900
<100, >900
Reynolds # Displays the Reynolds Number.
N/A
N/A
K(RE)
Displays the K Factor, based on the
Reynolds Number.
N/A
N/A
Raw Up
Displays raw upstream signal.
N/A
N/A
Raw Down
Displays raw downstream signal.
N/A
N/A